Remote Commands
Remote Command Listing
5
5-9
*OPC?
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Operations Complete query) Returns a 1 after all pending operations are
complete. This command causes program execution to pause until operations are
complete. (See *WAI.)
Response:
1
(all operations are complete)
Example:
*OPC?
returns
1
Returns
1
when all pending operations are complete.
OPER
Overlapped
IEEE-488
RS-232
Sequential
x
x
x
x
(Operate command) Activates the Calibrator output if it is in standby. This is the
same as pressing the Calibrator front panel
A
key.
Parameter:
(None)
Example:
OPER
Connect the selected output to the Calibrator front panel terminals. Also indicates
OPR on the display.
OPER?
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Operate query) Returns the operate/standby setting.
Response:
1
(Operate)
0
(Standby)
Example:
OPER?
returns
1
Returns
1
when the Calibrator is in operate.
*OPT?
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Options command) Returns a list of the installed hardware and software options.
Responses:
<option string>
,
<option string>
,
(options list, separated by
commas)
0
(no options are installed)
Example:
*OPT?
Returns
0
This command is reserved for future use.