First apply a four wire short to the 525A 4W RTD Ohm input. Set the 525A to the
400 ohm range. Note the reading. Set the 525A to the 4000 ohm range. Note the
reading. Move the 525A 4W RTD Ohm input to the output of the 5520A. Output
the same nominal values and note the 525A readings in the 525A reading column.
Subtract the actual 5520A value from the 525A reading and ensure that it is within
the specified tolerance.
5520 4W Ohms
Function
525A
INPUT
OUTPUT
HI
VOLTS
mA
RTD
HI
LO
HI
LO
LO
TC
SENSE
INPUT/OUTPUT
4W RTD
SET
RECALL
LOCAL
EXP
AUTOSET
INPUT
OUTPUT
ZERO
RNG LOCK
C / F
CJC
SETUP
20V PK
MAX
100V MAX
100mA MAX
20V PK
MAX
20V PK
MAX
525A
TEMPERATURE / PRESSURE CALIBRATOR
0
•
1
2
3
4
5
6
7
8
9
/
VOLTS
mA
TYPE
UNITS
TC
RTD
SHIFT
ENTER
STBY
OPR
CE
ajr27f.eps
Figure 6-8. Connection for Measuring Ohms
Testing Pressure Modules
The Fluke 700 series pressure modules are calibrated separately from the 525A.
The calibration follows the pressure module, so only a performance test is needed.
Connect any of the Fluke 700 series pressure modules to the pressure module
connector. Verify that the 525A reads pressure.