![Fluke 525A User Manual Download Page 116](http://html.mh-extra.com/html/fluke/525a/525a_user-manual_2309641116.webp)
525A
Users Manual
7-2
DC Voltage Specifications, Output
Absolute Uncertainty,
tcal
±
5
°
C
±
(ppm of
µ
V)
Stability
Ranges
1
90 days
1 year
24 hours,
±
1
°
C
±
(ppm of
µ
V)
Resolution
Maximum
Burden
2
0 to 100.000 mV
25
3
30
3
5 +2
1
µ
V
10 mA
0 to 1.00000 V
25
20
30
20
4 +20
10
µ
V
10 mA
0 to 10.0000 V
25
200
30
200
4 +200
100
µ
V
10 mA
0 to 100.000 V
25
2 mV
30
2 mV
5 +1 mV
1 mV
1 mA
TC Output
–
10 to 75.000 mV
25
3
µ
V
30
3
µ
V
5 +2
µ
V
1
µ
V
10
Ω
1. All outputs are positive only.
2. Remote sensing is not provided. Output resistance is < 1
Ω
.
Noise
Ranges
Bandwidth 0.1 to 10 Hz
(ppm of
µ
V p-p)
Bandwidth10 Hz to 10 kHz
(
µ
V rms)
0 to 100.000 mV
1
µ
V
6
µ
V
0 to 1.00000 V
10
µ
V
60
µ
V
0 to 10.0000 V
100
µ
V
600
µ
V
0 to 100.000 V
10 ppm+1 mV
20 mV
Summary of Contents for 525A