![Fluke 525A User Manual Download Page 89](http://html.mh-extra.com/html/fluke/525a/525a_user-manual_2309641089.webp)
Remote Commands
Remote Command Listing
5
5-17
TC_REF
Overlapped
IEEE-488
RS-232
Sequential
x
x
x
x
(Thermocouple Reference command) Sets whether the internal temperature sensor
(INT) or an external reference value (EXT) is used for Thermocouple (TC) outputs
and measurements. The Calibrator retains the TC reference setting until power off
or reset.
Parameters:
INT
EXT
Example:
TC_REF EXT
Set the thermocouple reference to external.
TC_REF?
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Thermocouple Reference query) Returns the source of the temperature being
used as a reference for thermocouple simulation and measurement. The choices are
Internal reference (INT) or External reference (EXT).
Responses:
INT
EXT
Example:
TC_REF?
returns
INT
Returns Internal.
Summary of Contents for 525A
Page 8: ...525A Users Manual vi ...