4180ETRS
Operators Manual
30
PROG:OPT:SETT n
Read or set the program settle option, 0 (apply default limit) or 1 (apply STABLE LIMIT setting).
Read Example: PROG:OPT:SETT?
Response: 1
Set Example: PROG:OPT:SETT 1
PROG:OPT:SOAK n
Read or set the program soak time, 0 to 500 minutes. Default 1 Read Example: PROG:OPT:SOAK?
Response: 15
Set Example: PROG:OPT:SOAK 350
PROG[n]:PAR? PAR
Read or set a program parameter, for a given program identified by n, n = 1 to 8, where par is the
parameter that you want to read. Parameters are SPOi, POIN, IRTE, DIST, and APER.
SPOi = set-points (1-8), where i =the value of one setpoint. POIN = the number of set-points for the
indicated program. IRTE= the emissivity
ε
value, default 0.928.
DIST= the distance from the target to the DUT in cm, 0.1 to 999.9.
APER= yes or no to prompt user for the aperture. 0 = none, 1 = prompt user
Read Example: PROG[1]PAR? DIST?
Response: 100.0
Set Example: PROG[1]:PAR DIST,150
PROG:PAR:CAT?
Read a list of program parameters.
Read Example: PROG:OPT:SOAK?
Response: SPO8, POIN, IRTE, DIST, APER
PROG:PROM:ADV
Advance to the next program step if waiting for user input.
Set Example: PROG:PROM:ADV
PROG:PROM:STAT?
Read the manual program advance prompt state, 0(operating or program off) or 1 (waiting for user
input).
Read Example: PROG:STAT? Response: 0