35
SOUR:SENS:DATA?
Reads the apparent temperature, in °C or °F.
Read Example: SOUR:SENS:DATA?
Response: 24.091 (current apparent temperature)
SOUR:SPO[?] [n]
Set the control set-point, °C or °F, where "n" is a real value with acceptance limits based on the model.
Read Example: SOUR:SPO?
Response: 50.000
Set Example: SOUR:SPO 37.00
SOUR:STAB:BEEP[?] [n]
Read or set the stability alert (beep) enable where "n" is a value 0 or 1. [0] is disable, [1] is enable
beep. Default: 1 (Enable Beep)
Read Example: SOUR:STAB:BEEP? Response: 1
Set Example: SOUR:STAB:BEEP 0 Enable or disable the audible stability alert.
SOUR:STAB:DAT?
Read the control temperature stability, °C or °F
Example: SOUR:STAB:DAT?
Response: 0.306
The controller stability is returned.
SOUR:STAB:LIM[?] [n]
Read or set the control temperature stability limit, °C or °F where "n" is a positive real value.
Range = {0.01 to 5.0 (°C)}; Default: 0.1 (°C)
Read Example: SOUR:STAB:LIM? Response: 0.1
Set Example: SOUR:STAB:LIM 0.09
Parameter
Min
Max
Default
4180ETRS
10
50
25.00