Precision Infrared Calibrator
Digital Communication Interface
29
OUTP:DATA?
Read the main heat output percent
Example: OUTP1:DATA?
Response: 18.0
This command returns the current main zone heater duty cycle.
OUTP:STAT[?] [0|1]
Read or set the Main Heat output enable, off [0] or on [1]
Read Example: OUTP:STAT?
Response: 0
Set Example: OUTP:STAT 1
This command reads or sets the active heating or cooling output status. 0 is returned if the output
status is off, and 1 is returned if the output status is on.
PROG:[n]NAME?
Read or set the program name by identifier by n,1 - 8. Where name consists of 1 to 10
Characters 0 to 9, A to Z and -. If n is omitted, the selected program is assumed.
Read Example: PROG:[1]NAME?
Response: PROG
Set Example: PROG:[1]NAME newprog01A
PROG:OPT:ADV n
Read or set the program advance option, 0 (prompt) or 1 (continue automatically)
Read Example: PROG:OPT:ADV?
Response: 0
Set Example: PROG:OPT:ADV 1
PROG:OPT:CYCL n
Read or set the program cycles, 1 to 999, default 1.
Read Example: PROG:OPT:CYCL?
Response: 123
Set Example: PROG:OPT:CYCL 150