Product Overview and Specifications
Measurement Specifications
1
1-15
DC Voltage Additional Errors
Digits
NPLC
Ch. x01 – x20
Additional NPLC Noise
Error
6 ½
200
add 2
µ
V
-
6 ½
100
add 2
µ
V
-
6 ½
10 (Slow)
add 2
µ
V
-
5 ½
1 (Medium)
add 2
µ
V
add 0.0008 % of range
4 ½
0.2 (Fast)
-
add 0.002 % of range + 12
µ
V
4 ½
0.02
-
add 0.014 % of range + 17
µ
V
AC Voltage
AC voltage specifications are for ac sine wave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,
add an additional error of 0.1 % of range. For 50 kHz to 100 kHz, add 0.13 % of range.
Maximum Input
...................................................... 300 V rms or 425 V peak or 3 × 10
7
volts-Hertz product (whichever is
less) for any range.
Measurement Method
........................................... AC-coupled true-rms. Measures the ac component of input with up to
300 V dc bias on any range.
AC Filter Bandwidth:
Slow .................................................................... 20 Hz
Fast ..................................................................... 200 Hz
Maximum Crest Factor
......................................... 5:1 at full scale
Additional Crest Factor Errors
............................ Crest factor 1-2, 0.05 % of full scale
Crest factor 2-3, 0.2 % of full scale
Crest factor 3-4, 0.4 % of full scale
Crest factor 4-5, 0.5 % of full scale
AC Voltage Input Characteristics
Range Resolution
Resolution
Input Impedance
4 ½
Digits
5 ½
Digits
6 ½
Digits
100 mV
100.0000 mV
10
µ
V 1
µ
V 0.1
µ
V
1 M
Ω
±2 % shunted by 150 pF
1 V
1.000000 V
100
µ
V 10
µ
V 1
µ
V
10 V
10.00000 V
1 mV
100 µV
10
µ
V
100 V
100.0000 V
10 mV
1 mV
100
µ
V
300 V
300.000 V
100 mV
10 mV
1 mV
1.888.475.5235