Input and Channel Configuration
Channel Configuration
3
3-25
Set up a totalizer channel as follows (see Table 3-10):
1.
Connect the input to the TOT input terminal then insert it into the rear-panel Digital
I/O port.
2.
Set the channel to ON (see “Set Channels to ON or OFF” on page 3-10). When the
channel is set to ON, a channel configuration sequence shows on the display that
steps the user through an initial configuration sequence as shown in Table 3-10. If the
channel was previously configured, push
to cancel the configuration sequence
to use the last channel configuration.
3.
Use the information in Table 3-10 to configure the totalizer channel. Push
or
to
highlight the channel setting then push
to edit the setting.
4.
To see the totalizer count, monitor Ch402 in the Scan/Monitor menu.
Table 3-10. Totalizer Channel Configuration
1
2
3
hce054.eps
Item Function
Input a custom alpha-numeric label to help identify the channel (optional).
Read mode selection. See “Read Mode” on page 3-24.
Set the Debounce feature to ON or OFF. See “Debounce” on page 3-24.