![Fluke 2620A Manual Download Page 69](http://html1.mh-extra.com/html/fluke/2620a/2620a_manual_2309537069.webp)
Theory of Operation (2620A/2625A)
Detailed Circuit Description
2
2-21
Table 2-5. Function Relay States
Relay Position
Function
A3K17
A3K16
A3K15
DC mV, 3V,Thermocouples
Reset
Set
Set
DC 30V, 300V
Set
Set
Set
ACV
Set
Set
Reset
Ohms, RTDs
Reset
Reset
Set
Frequency
Set
Set
Reset
2-54. DC Volts and Thermocouples
For the 3V and lower ranges (including thermocouples), the HI input signal is applied
directly to the A3U8 analog processor through A3R11, A3K17, and A3R42. Capacitor
A3C27 filters this input, which the analog processor then routes through S2 and other
internal switches, through the passive filter, and to the internal a/d converter. The LO
SENSE signal is applied to A3U8 through A3R35 and routed through internal switch
A3U8-S19 to LO of the a/d converter.
Guard signals MGRD and RGRD are driven by an amplifier internal to A3U8 to a
voltage appropriate for preventing leakage from the input HI signal under high humidity
conditions.
For the 30V range, the HI signal is scaled by resistor network A3Z4. Here, the input is
applied to pin 1 of A3Z4 so that an approximate 100:1 divider is formed by the 10-M
Ω
and 100.5-k
Ω
resistors in A3Z4 when analog processor switches S3 and S13 are closed.
The attenuated HI input is then sent through internal switch S12 to the passive filter and
the a/d converter. Input LO is sensed through analog processor switch S18 and resistor
A3R34.
For the 300V range (Figure 2-4), the HI signal is again scaled by A3Z4. The input is
applied to pin 1 of A3Z4, and a 1000:1 divider is formed by the 10-M
Ω
and 10.01-k
Ω
resistors when switches S3 and S9 are closed in A3Z4. The attenuated HI input is then
sent through internal switch S10 to the passive filter and the a/d converter. LO is sensed
through analog processor switch S18 and resistor A3R34.
2-55. Ohms and RTDs
Resistance measurements are made using a ratio ohms technique, as shown in Figure 2-
5. A stable voltage source is connected in series with the reference resistor in A3Z4 and
the unknown resistor. Since the same current flows through both resistors, the unknown
resistance can be determined by multiplying the ratio of the voltage drops across the
reference and the unknown resistors by the known reference resistor value.
For the RTD, 300
Ω
, 3-k
Ω
, and 30-k
Ω
ranges, the ratio technique is implemented by
integrating the voltage across the unknown resistance for a fixed period of time and then
integrating the negative of the voltage across the reference resistance for a variable time
period. In this way, each minor cycle result gives the ratio directly.
For the 300-k
Ω
, 3-M
Ω
, and 10-M
Ω
ranges, the ratio is determined by performing two
separate voltage measurements in order to improve noise rejection. One fixed-period
integration is performed on the voltage across the unknown resistance, and the second
integration is performed on the voltage across the reference resistance. The ratio of the
two fixed-period voltge measurements is then computed by Microcontroller A3U9. The
resistance measurement result is determined when A3U9 multiplies the ratio by the
reference resistance value.
Summary of Contents for 2620A
Page 4: ......
Page 18: ...HYDRA Service Manual 1 2...
Page 48: ...HYDRA Service Manual 1 32...
Page 140: ...HYDRA Service Manual 3 16...
Page 234: ...HYDRA Service Manual 5A 34...
Page 236: ...HYDRA Service Manual 6 2...
Page 254: ...HYDRA Service Manual 6 20 2620A 1601 s63f eps Figure 6 3 2620A 2625A A1 Main PCA...
Page 258: ...HYDRA Service Manual 6 24 2635A 1601 s64f eps Figure 6 4 2635A A1 Main PCA...
Page 260: ...HYDRA Service Manual 6 26 2620A 4002 CKT 1 CKT 2 s65f eps Figure 6 5 A2 Display PCA...
Page 272: ...HYDRA Service Manual 6 38...
Page 274: ...HYDRA Service Manual 7 2...
Page 282: ...HYDRA Service Manual 7 10...
Page 284: ...HYDRA Service Manual 8 2...
Page 285: ...Schematic Diagrams 8 8 3 Figure 8 1 A1 Main PCA 2620A 2625A 2620A 1601 s88f eps...
Page 288: ...Schematic Diagrams 8 8 6 Figure 8 1 A1 Main PCA 2620A 2625A cont 2620A 1001 3 of 4 s73f eps...
Page 292: ...Schematic Diagrams 8 8 10 2635A 1001 2 of 5 Figure 8 2 A1 Main PCA 2635A cont s76f eps...
Page 293: ...Schematic Diagrams 8 8 11 2635A 1001 3 of 5 Figure 8 2 A1 Main PCA 2635A cont s77f eps...
Page 296: ...Schematic Diagrams 8 8 14 2620A 4002 CKT 1 CKT 2 Figure 8 3 A2 Display PCA s90f eps...
Page 300: ...Schematic Diagrams 8 8 18 2620A 1003 2 of 3 Figure 8 4 A3 A D Converter PCA cont s82c eps...
Page 301: ...Schematic Diagrams 8 8 19 2620A 1003 3 of 3 Figure 8 4 A3 A D Converter PCA cont s83c eps...
Page 302: ...Schematic Diagrams 8 8 20 2620A 1604 Figure 8 5 A4 Analog Input PCA s92f eps...
Page 304: ...Schematic Diagrams 8 8 22 2620A 1605 Figure 8 6 A5 IEEE 488 Interface PCA 2620A Only s93f eps...
Page 306: ...Schematic Diagrams 8 8 24 2625A 1606 Figure 8 7 A6 Memory PCA 2625A s94f eps...
Page 310: ...Schematic Diagrams 8 8 28...