![Fluke 2620A Manual Download Page 113](http://html1.mh-extra.com/html/fluke/2620a/2620a_manual_2309537113.webp)
Theory of Operation (2635A)
Detailed Circuit Description
2A
2A-31
BUFFER
A3Z2
INTEGRATOR
INTEGRATE
REFERENCE
INTEGRATE
INPUT
–REFERENCE
(+ INPUT)
REFERENCE
+ REFERENCE
(– INPUT)
S77
A3C13
+
_
_
+
+
_
A/D
COMPARATOR
+
_
COUNTER
INPUT HI
INPUT LO
s17f.eps
Figure 2A-7. A/D Converter Simplified Schematic (2635A)
In both the slow and fast measurement rates, the a/d converter uses its ±300 mV range
for most measurement functions and ranges. The primary exceptions are that the 3V dc
range is measured on the a/d converter 3V range, thermocouples are measured on the
±100 mV range, and the temperature reference is measured on the 1V a/d converter
range. The typical overload point on a slow rate 30000 count range is 32000 display
counts; the typical overload point on a fast rate 3000 count range is 3200 display counts.
During the integrate phase, the a/d buffer in the A3U8 Analog Measurement Processor
applies the signal to be measured to one of the four integrator input resistors in network
A3Z2. As shown on the A/D Converter schematic diagram in Section 8, the choice of
resistor selects the a/d converter range. Switch S69 connects the buffer output through
pin B.1 for the 100-mV range, S71 connects the output through B.32 for the 300 mV
range, S73 connects to pin B1 for the 1V range, and S75 sets up the 3V range through
pin B3.2.
The current through the selected integrator input resistor charges integrator capacitor
A3C13, with the current dependent on the buffer output voltage. After the integrate
phase, the buffer is connected to the opposite polarity reference voltage, and the
integrator integrates back toward zero capacitor voltage until the comparator trips. An
internal counter measures this variable integrate time. If the a/d converter input voltage
is too high, the integrator overloads and does not return to its starting point by the end of
the measurement phase. Switch S77 is then turned on to discharge integrate capacitor
A3C13.
The reference voltage used during the variable integrate period for voltage (and high
ohms) conversions is generated from zener reference diode A3VR1, which is time and
temperature stable. The reference amplifier in the Analog Measurement Processor, along
with resistors A3R15, A3R18, and A3R21, pulls approximately 2 mA of current through
the zener. Resistors in network A3Z2 divide the zener voltage down to the reference
1.05V required by the A/D Converter.
Summary of Contents for 2620A
Page 4: ......
Page 18: ...HYDRA Service Manual 1 2...
Page 48: ...HYDRA Service Manual 1 32...
Page 140: ...HYDRA Service Manual 3 16...
Page 234: ...HYDRA Service Manual 5A 34...
Page 236: ...HYDRA Service Manual 6 2...
Page 254: ...HYDRA Service Manual 6 20 2620A 1601 s63f eps Figure 6 3 2620A 2625A A1 Main PCA...
Page 258: ...HYDRA Service Manual 6 24 2635A 1601 s64f eps Figure 6 4 2635A A1 Main PCA...
Page 260: ...HYDRA Service Manual 6 26 2620A 4002 CKT 1 CKT 2 s65f eps Figure 6 5 A2 Display PCA...
Page 272: ...HYDRA Service Manual 6 38...
Page 274: ...HYDRA Service Manual 7 2...
Page 282: ...HYDRA Service Manual 7 10...
Page 284: ...HYDRA Service Manual 8 2...
Page 285: ...Schematic Diagrams 8 8 3 Figure 8 1 A1 Main PCA 2620A 2625A 2620A 1601 s88f eps...
Page 288: ...Schematic Diagrams 8 8 6 Figure 8 1 A1 Main PCA 2620A 2625A cont 2620A 1001 3 of 4 s73f eps...
Page 292: ...Schematic Diagrams 8 8 10 2635A 1001 2 of 5 Figure 8 2 A1 Main PCA 2635A cont s76f eps...
Page 293: ...Schematic Diagrams 8 8 11 2635A 1001 3 of 5 Figure 8 2 A1 Main PCA 2635A cont s77f eps...
Page 296: ...Schematic Diagrams 8 8 14 2620A 4002 CKT 1 CKT 2 Figure 8 3 A2 Display PCA s90f eps...
Page 300: ...Schematic Diagrams 8 8 18 2620A 1003 2 of 3 Figure 8 4 A3 A D Converter PCA cont s82c eps...
Page 301: ...Schematic Diagrams 8 8 19 2620A 1003 3 of 3 Figure 8 4 A3 A D Converter PCA cont s83c eps...
Page 302: ...Schematic Diagrams 8 8 20 2620A 1604 Figure 8 5 A4 Analog Input PCA s92f eps...
Page 304: ...Schematic Diagrams 8 8 22 2620A 1605 Figure 8 6 A5 IEEE 488 Interface PCA 2620A Only s93f eps...
Page 306: ...Schematic Diagrams 8 8 24 2625A 1606 Figure 8 7 A6 Memory PCA 2625A s94f eps...
Page 310: ...Schematic Diagrams 8 8 28...