![Fluke 2620A Manual Download Page 198](http://html1.mh-extra.com/html/fluke/2620a/2620a_manual_2309537198.webp)
HYDRA
Service Manual
5-28
5-23. Retrieving Calibration Constants
If a calibration error is suspected, the stored constant can be retrieved and verified over
the computer interface. Acceptable calibration constants for each function and range are
listed in Table 5-9 (software version 5.4 and higher) or 5-10 (software versions lower
than 5.4.) Retrieve the constant with the following command:
CAL_CONST? xx
(where xx denotes the calibration constant number)
Table 5-10. Calibration Faults (for sotware versions lower than 5.4)
Input
Range
Calibration Constant
Related Components
Number
Acceptable Values
DC Volts
0.09000V
0.9000V
0.29000V
2.9000V
29.000V
290.00V
100 mV
1V
30 mV0
3V
30V
300V
1
2
3
4
5
6
1.0315 to 1.1565
1.0340 to 1.1540
1.0315 to 1.1565
1.0315 to 1.1565
1.0340 to 1.1640
1.0290 to 1.1590
A3VR1, A3Z2
A3VR1, A3Z2
A3VR1, A3Z2
A3VR1, A3Z2
A3VR1, A3Z2, A3Z4
A3VR1, A3Z2, A3Z4
AC Volts (1 kHz)
0.02900V
0.29000V
0.2900V
2.9000V
29.000V
290.00V
300 mV
300 mV
3V
3V
30V
300V
7
8
9
10
11
12
-0.001 to 0.001
1.0040 to 1.1840
-0.01 to 0.01
1.0040 to 1.1840
1.0040 to 1.1840
1.0040 to 1.1840
A3U6, A3VR1, A3Z1, A3Z2, A3Z3
A3U6, A3VR1, A3Z1, A3Z2, A3Z3
A3U6, A3VR1, A3Z1, A3Z2, A3Z3
A3U6, A3VR1, A3Z1, A3Z2, A3Z3
A3U6, A3VR1, A3Z1, A3Z2, A3Z3
A3U6, A3VR1, A3Z1, A3Z2, A3Z3
Ohms
290.00
Ω
2.9000 k
Ω
29.000 k
Ω
v
290.00 k
Ω
2.9000 M
Ω
2.9000 M
Ω
300
Ω
3 k
Ω
30 k
Ω
300 k
Ω
3 M
Ω
10 M
Ω
13
14
15
16
17
18
0.9965 to 1.0115
0.9975 to 1.0125
1.0015 to 1.0165
0.9965 to 1.0115
0.9990 to 1.0090
0.9990 to 1.0090
A3Z2, A3Z4
A3Z2, A3Z4
A3Z2, A3Z4
A3Z2, A3Z4
A3Z2, A3Z4
A3Z2, A3Z4
Frequency
10.000 kHz
2.9V rms
19
0.9995 to 1.0005
A3Y2
5-24. Replacing the EEPROM (A1U1)
The EEPROM provides nonvolatile storage for the instrument serial number, some of the
instrument configuration, and all calibration information. If the EEPROM is replaced
during repair, the new EEPROM can be programmed with the 7-digit serial number
found on the rear panel of the instrument or any 7-digit identifier of your choosing. Note
that the serial number is not programmed prior to shipment from the factory.
Summary of Contents for 2620A
Page 4: ......
Page 18: ...HYDRA Service Manual 1 2...
Page 48: ...HYDRA Service Manual 1 32...
Page 140: ...HYDRA Service Manual 3 16...
Page 234: ...HYDRA Service Manual 5A 34...
Page 236: ...HYDRA Service Manual 6 2...
Page 254: ...HYDRA Service Manual 6 20 2620A 1601 s63f eps Figure 6 3 2620A 2625A A1 Main PCA...
Page 258: ...HYDRA Service Manual 6 24 2635A 1601 s64f eps Figure 6 4 2635A A1 Main PCA...
Page 260: ...HYDRA Service Manual 6 26 2620A 4002 CKT 1 CKT 2 s65f eps Figure 6 5 A2 Display PCA...
Page 272: ...HYDRA Service Manual 6 38...
Page 274: ...HYDRA Service Manual 7 2...
Page 282: ...HYDRA Service Manual 7 10...
Page 284: ...HYDRA Service Manual 8 2...
Page 285: ...Schematic Diagrams 8 8 3 Figure 8 1 A1 Main PCA 2620A 2625A 2620A 1601 s88f eps...
Page 288: ...Schematic Diagrams 8 8 6 Figure 8 1 A1 Main PCA 2620A 2625A cont 2620A 1001 3 of 4 s73f eps...
Page 292: ...Schematic Diagrams 8 8 10 2635A 1001 2 of 5 Figure 8 2 A1 Main PCA 2635A cont s76f eps...
Page 293: ...Schematic Diagrams 8 8 11 2635A 1001 3 of 5 Figure 8 2 A1 Main PCA 2635A cont s77f eps...
Page 296: ...Schematic Diagrams 8 8 14 2620A 4002 CKT 1 CKT 2 Figure 8 3 A2 Display PCA s90f eps...
Page 300: ...Schematic Diagrams 8 8 18 2620A 1003 2 of 3 Figure 8 4 A3 A D Converter PCA cont s82c eps...
Page 301: ...Schematic Diagrams 8 8 19 2620A 1003 3 of 3 Figure 8 4 A3 A D Converter PCA cont s83c eps...
Page 302: ...Schematic Diagrams 8 8 20 2620A 1604 Figure 8 5 A4 Analog Input PCA s92f eps...
Page 304: ...Schematic Diagrams 8 8 22 2620A 1605 Figure 8 6 A5 IEEE 488 Interface PCA 2620A Only s93f eps...
Page 306: ...Schematic Diagrams 8 8 24 2625A 1606 Figure 8 7 A6 Memory PCA 2625A s94f eps...
Page 310: ...Schematic Diagrams 8 8 28...