![Fluke 2620A Manual Download Page 164](http://html1.mh-extra.com/html/fluke/2620a/2620a_manual_2309537164.webp)
HYDRA
Service Manual
4-24
Note
The 300 k
Ω
, 3 M
Ω
, and 10 M
Ω
ranges are sensitive to noise. Any
movement of the input leads can cause noisy readings. Use shielded leads
and verify these two calibration points at the conclusion of calibration.
Table 4-10. AC Volts Calibration
Command
Response
Action
CAL 2
=>
Puts Hydra in VAC Calibration.
CAL_REF?
+029.00E-3
You output 29 mV ac at 1 kHz from the 5700A. Wait about 8 seconds.
CAL_STEP?
Hydra computes calibration constant 7 and returns the calibrated reading
(for example, +029.00E-3.)
Note
If the input is incorrect, the "!>" response signifies that a Device
Dependent Error was generated. The calibration step could not be
executed. Verify that the input to Hydra channel 1 is the correct value and
polarity. Also verify that the 5700A is in OPERATE mode. If the input is
correct, Hydra may require repair.
CAL_REF?
+290.00E-3
You output 290 mV at 1 kHz from the 5700A. Wait 8 seconds.
CAL_STEP?
Hydra computes calibration constant 8 and returns the calibrated reading.
CAL_REF?
+0.2900E+0
You output 290 mV at 1 kHz from the 5700A and wait 8 seconds.
CAL_STEP?
Hydra computes calibration constant 9 and returns the calibrated reading.
CAL_REF?
+2.9000E+0
You output 2.9V at 1 kHz from the 5700A and wait 8 seconds.
CAL_STEP?
Hydra computes calibration constants 10 and 11 and returns the
calibrated reading. (For software versions lower than 5.4, this step
computes calibration constant 10 only.)
CAL_REF?
+29.000E+0
You output 29V at 1 kHz and wait 8 seconds.
CAL_STEP?
Hydra computes calibration constants 12 and 13 and returns the
calibrated reading. (For software versions lower than 5.4, this step
computes calibration constant 11 only.)
CAL_REF?
+290.00E+0
You output 290V at 1 kHz and wait 8 seconds.
CAL_STEP?
Hydra computes calibration constant 14 and returns the calibrated
reading. (For software versions lower than 5.4, this step computes
calibration constant 12.)
4-28. Reference Junction Calibration
Note
This procedure is necessary only if the Input Module has been repaired or
damaged, or if R3 on the Input Module has been inadvertently adjusted.
If thermocouple readings taken in the Thermocouple Temperature Accuracy Test section
of the performance tests are found to be out of tolerance, the Input Module Reference
Junction may require calibration. First, check the volts dc calibration. If volts dc
calibration is correct, perform the following steps:
1.
Perform the Volts DC group calibration.
2.
Switch OFF power to Hydra, and remove the Input Module from the rear of the
instrument.
Summary of Contents for 2620A
Page 4: ......
Page 18: ...HYDRA Service Manual 1 2...
Page 48: ...HYDRA Service Manual 1 32...
Page 140: ...HYDRA Service Manual 3 16...
Page 234: ...HYDRA Service Manual 5A 34...
Page 236: ...HYDRA Service Manual 6 2...
Page 254: ...HYDRA Service Manual 6 20 2620A 1601 s63f eps Figure 6 3 2620A 2625A A1 Main PCA...
Page 258: ...HYDRA Service Manual 6 24 2635A 1601 s64f eps Figure 6 4 2635A A1 Main PCA...
Page 260: ...HYDRA Service Manual 6 26 2620A 4002 CKT 1 CKT 2 s65f eps Figure 6 5 A2 Display PCA...
Page 272: ...HYDRA Service Manual 6 38...
Page 274: ...HYDRA Service Manual 7 2...
Page 282: ...HYDRA Service Manual 7 10...
Page 284: ...HYDRA Service Manual 8 2...
Page 285: ...Schematic Diagrams 8 8 3 Figure 8 1 A1 Main PCA 2620A 2625A 2620A 1601 s88f eps...
Page 288: ...Schematic Diagrams 8 8 6 Figure 8 1 A1 Main PCA 2620A 2625A cont 2620A 1001 3 of 4 s73f eps...
Page 292: ...Schematic Diagrams 8 8 10 2635A 1001 2 of 5 Figure 8 2 A1 Main PCA 2635A cont s76f eps...
Page 293: ...Schematic Diagrams 8 8 11 2635A 1001 3 of 5 Figure 8 2 A1 Main PCA 2635A cont s77f eps...
Page 296: ...Schematic Diagrams 8 8 14 2620A 4002 CKT 1 CKT 2 Figure 8 3 A2 Display PCA s90f eps...
Page 300: ...Schematic Diagrams 8 8 18 2620A 1003 2 of 3 Figure 8 4 A3 A D Converter PCA cont s82c eps...
Page 301: ...Schematic Diagrams 8 8 19 2620A 1003 3 of 3 Figure 8 4 A3 A D Converter PCA cont s83c eps...
Page 302: ...Schematic Diagrams 8 8 20 2620A 1604 Figure 8 5 A4 Analog Input PCA s92f eps...
Page 304: ...Schematic Diagrams 8 8 22 2620A 1605 Figure 8 6 A5 IEEE 488 Interface PCA 2620A Only s93f eps...
Page 306: ...Schematic Diagrams 8 8 24 2625A 1606 Figure 8 7 A6 Memory PCA 2625A s94f eps...
Page 310: ...Schematic Diagrams 8 8 28...