![Fluke 2620A Manual Download Page 151](http://html1.mh-extra.com/html/fluke/2620a/2620a_manual_2309537151.webp)
Performance Testing and Calibration
Performance Tests
4
4-11
4-9.
Open Thermocouple Response Test
Use the following procedure to test the open thermocouple response:
1.
Switch OFF power to the instrument and disconnect all high voltage inputs.
2.
Remove the Input Module from the rear of the instrument. Open the Input Module
and connect test leads to the H (high) and L (low) terminals of channel 1. Reinstall
the Input Module into the instrument.
3.
Reconnect power and switch the instrument ON.
4.
Connect the test leads from the Input Module to an 820 ohm resistor.
5.
Select the temperature function and K thermocouple type for channel 1. Then press
MON.
6.
The value displayed should approximate the ambient temperature.
7.
Replace the 820-ohm resistor with a 4-kilohm resistor to simulate a high resistance
or open thermocouple.
8.
Verify a reading of "otc".
9.
The Open Thermocouple Response Test is complete. However if you desire to
perform this test on any other Input Module channel (2 through 20) repeat steps 1
through 8, substituting the appropriate channel number.
4-10. RTD Temperature Accuracy Test
The following two RTD Temperature Accuracy Tests are different in that one uses a
Decade Resistance Source and the other uses an RTD. Only one of the tests needs to be
performed to verify operation.
4-11. RTD Temperature Accuracy Test (Using Decade Resistance Source).
Verify that the channel 0 accuracy verification tests for dc volts and 300-ohm range meet
minimum acceptable levels.
1.
Switch OFF power to the instrument and disconnect all high voltage inputs.
2.
Remove the Input Module from the rear of the instrument. Open the Input Module
and connect a pair of test leads (keep as short as possible) to the H (high) and L
(low) terminals of channel 1. For 4-terminal performance testing, connect a second
pair of test leads to the H (high) and L (low) terminals of channel 11. Reinstall the
Input Module into the instrument.
3.
Connect the channel 1 test leads to the Output HI and LO terminals on the Decade
Resistance Source. For 4-terminal performance testing, also connect channel 11’s
test leads to the Output HI and LO terminals of the Decade Resistance Source.
Connect as shown in Figures 4-1 and 4-2.
Note
4-terminal connections are made using pairs of channels. 4-terminal
measurements can only be made on channels 1 through 10. The
accompanying pairs are channels 11 through 20.
4.
Switch the instrument ON.
5.
Select the 4-terminal RTD temperature function, RTD type PT, for channel 1 on
Hydra.
6.
Press MON. For each resistance, ensure that the display reads within the range
shown in Table 4-5.
Summary of Contents for 2620A
Page 4: ......
Page 18: ...HYDRA Service Manual 1 2...
Page 48: ...HYDRA Service Manual 1 32...
Page 140: ...HYDRA Service Manual 3 16...
Page 234: ...HYDRA Service Manual 5A 34...
Page 236: ...HYDRA Service Manual 6 2...
Page 254: ...HYDRA Service Manual 6 20 2620A 1601 s63f eps Figure 6 3 2620A 2625A A1 Main PCA...
Page 258: ...HYDRA Service Manual 6 24 2635A 1601 s64f eps Figure 6 4 2635A A1 Main PCA...
Page 260: ...HYDRA Service Manual 6 26 2620A 4002 CKT 1 CKT 2 s65f eps Figure 6 5 A2 Display PCA...
Page 272: ...HYDRA Service Manual 6 38...
Page 274: ...HYDRA Service Manual 7 2...
Page 282: ...HYDRA Service Manual 7 10...
Page 284: ...HYDRA Service Manual 8 2...
Page 285: ...Schematic Diagrams 8 8 3 Figure 8 1 A1 Main PCA 2620A 2625A 2620A 1601 s88f eps...
Page 288: ...Schematic Diagrams 8 8 6 Figure 8 1 A1 Main PCA 2620A 2625A cont 2620A 1001 3 of 4 s73f eps...
Page 292: ...Schematic Diagrams 8 8 10 2635A 1001 2 of 5 Figure 8 2 A1 Main PCA 2635A cont s76f eps...
Page 293: ...Schematic Diagrams 8 8 11 2635A 1001 3 of 5 Figure 8 2 A1 Main PCA 2635A cont s77f eps...
Page 296: ...Schematic Diagrams 8 8 14 2620A 4002 CKT 1 CKT 2 Figure 8 3 A2 Display PCA s90f eps...
Page 300: ...Schematic Diagrams 8 8 18 2620A 1003 2 of 3 Figure 8 4 A3 A D Converter PCA cont s82c eps...
Page 301: ...Schematic Diagrams 8 8 19 2620A 1003 3 of 3 Figure 8 4 A3 A D Converter PCA cont s83c eps...
Page 302: ...Schematic Diagrams 8 8 20 2620A 1604 Figure 8 5 A4 Analog Input PCA s92f eps...
Page 304: ...Schematic Diagrams 8 8 22 2620A 1605 Figure 8 6 A5 IEEE 488 Interface PCA 2620A Only s93f eps...
Page 306: ...Schematic Diagrams 8 8 24 2625A 1606 Figure 8 7 A6 Memory PCA 2625A s94f eps...
Page 310: ...Schematic Diagrams 8 8 28...