Theory of Operation (2635A)
Detailed Circuit Description
2A
2A-25
•
The open thermocouple detect circuitry is protected against voltagetransient damage
by A3Q14 and A3Q15.
•
When measuring ac volts, the ac buffer is protected by dual-diodeclamp A3CR1 and
resistor network A3Z3.
•
Switching induced transients are also clamped by dual-diodeA3CR4 and capacitor
A3C33, and limited by resistor A3R33.
2A-51. Input Signal Conditioning
Each input is conditioned and/or scaled to a dc voltage appropriate for measurement by
the a/d converter. DC voltage applied to the a/d converter can be handled on internal
ranges of 0.1V, 0.3V, 1V, or 3V. Therefore, high-voltage dc inputs are scaled, and ohms
inputs are converted to a dc voltage. Line voltage level ac inputs are first scaled and then
converted to a dc voltage. Noise rejection is provided by passive and active filters.
2A-52. Function Relays
Latching relays A3K15, A3K16, and A3K17 route the input signal to the proper circuit
blocks to implement the desired measurement function. These relays are switched when
a 6-millisecond pulse is applied to the appropriate reset or set coil by the NPN
Darlington drivers in IC A3U10. The A/D Microcontroller A3U9 controls the relay drive
pulses by setting the outputs of port 6. Since the other end of the relay coil is connected
to the VDDR supply, a magnetic field is generated, causing the relay armature and
contacts to move to (or remain in) the desired position. Function relay states are defined
in Table 2A-5.
Table 2A-5. Function Relay States (2635A)
Relay Position
Function
A3K17
A3K16
A3K15
DC mV, 3V,Thermocouples
Reset
Set
Set
DC 30V, 300V
Set
Set
Set
ACV
Set
Set
Reset
Ohms, RTDs
Reset
Reset
Set
Frequency
Set
Set
Reset
2A-53. DC Volts and Thermocouples
For the 3V and lower ranges (including thermocouples), the HI input signal is applied
directly to the A3U8 analog processor through A3R11, A3K17, and A3R42. Capacitor
A3C27 filters this input, which the analog processor then routes through S2 and other
internal switches, through the passive filter, and to the internal a/d converter. The LO
SENSE signal is applied to A3U8 through A3R35 and routed through internal switch
A3U8-S19 to LO of the a/d converter.
Guard signals MGRD and RGRD are driven by an amplifier internal to A3U8 to a
voltage appropriate for preventing leakage from the input HI signal under high humidity
conditions.
For the 30V range, the HI signal is scaled by resistor network A3Z4. Here, the input is
applied to pin 1 of A3Z4 so that an approximate 100:1 divider is formed by the 10-M
Ω
and 100.5-k
Ω
resistors in A3Z4 when analog processor switches S3 and S13 are closed.
The attenuated HI input is then sent through internal switch S12 to the passive filter and
the a/d converter. Input LO is sensed through analog processor switch S18 and resistor
A3R34.
Summary of Contents for 2620A
Page 4: ......
Page 18: ...HYDRA Service Manual 1 2...
Page 48: ...HYDRA Service Manual 1 32...
Page 140: ...HYDRA Service Manual 3 16...
Page 234: ...HYDRA Service Manual 5A 34...
Page 236: ...HYDRA Service Manual 6 2...
Page 254: ...HYDRA Service Manual 6 20 2620A 1601 s63f eps Figure 6 3 2620A 2625A A1 Main PCA...
Page 258: ...HYDRA Service Manual 6 24 2635A 1601 s64f eps Figure 6 4 2635A A1 Main PCA...
Page 260: ...HYDRA Service Manual 6 26 2620A 4002 CKT 1 CKT 2 s65f eps Figure 6 5 A2 Display PCA...
Page 272: ...HYDRA Service Manual 6 38...
Page 274: ...HYDRA Service Manual 7 2...
Page 282: ...HYDRA Service Manual 7 10...
Page 284: ...HYDRA Service Manual 8 2...
Page 285: ...Schematic Diagrams 8 8 3 Figure 8 1 A1 Main PCA 2620A 2625A 2620A 1601 s88f eps...
Page 288: ...Schematic Diagrams 8 8 6 Figure 8 1 A1 Main PCA 2620A 2625A cont 2620A 1001 3 of 4 s73f eps...
Page 292: ...Schematic Diagrams 8 8 10 2635A 1001 2 of 5 Figure 8 2 A1 Main PCA 2635A cont s76f eps...
Page 293: ...Schematic Diagrams 8 8 11 2635A 1001 3 of 5 Figure 8 2 A1 Main PCA 2635A cont s77f eps...
Page 296: ...Schematic Diagrams 8 8 14 2620A 4002 CKT 1 CKT 2 Figure 8 3 A2 Display PCA s90f eps...
Page 300: ...Schematic Diagrams 8 8 18 2620A 1003 2 of 3 Figure 8 4 A3 A D Converter PCA cont s82c eps...
Page 301: ...Schematic Diagrams 8 8 19 2620A 1003 3 of 3 Figure 8 4 A3 A D Converter PCA cont s83c eps...
Page 302: ...Schematic Diagrams 8 8 20 2620A 1604 Figure 8 5 A4 Analog Input PCA s92f eps...
Page 304: ...Schematic Diagrams 8 8 22 2620A 1605 Figure 8 6 A5 IEEE 488 Interface PCA 2620A Only s93f eps...
Page 306: ...Schematic Diagrams 8 8 24 2625A 1606 Figure 8 7 A6 Memory PCA 2625A s94f eps...
Page 310: ...Schematic Diagrams 8 8 28...