![Fluke 2620A Manual Download Page 188](http://html1.mh-extra.com/html/fluke/2620a/2620a_manual_2309537188.webp)
HYDRA
Service Manual
5-18
Table 5-5. AC Volts HI Troubleshooting
Checkpoint
Signal Description
Possible Fault
A3R11 HI
Input
A3K1 through A3K14, A3U4, A3U5, A3U11, A3U12
A3L1, A3L2, A3L3
A3Z3 pin 1
Input
A3R11, A3C31, A3K15
A3U6 pin 13
Amplified (X 2.5) input
A3U7, A3Z3, A3Q3 through A3Q9, A3C15, A3C16,
A3R24, A3A25, A3R26, A3R27, A3R28, A3C23,
A3U6, A3Q13, A3U8
A3Z1 pin 2
DC equivalent of original input
A3Z1, A3U8, A3R20, A3C6, A3C7, A3C10, A3R16,
A3R17
A3U8 pin 61
DC equivalent of original input
A3Z1, A3U8, A3R20, A3C6, A3C7, A3C10, A3R16,
A3R17
5-13. Ohms Troubleshooting
Setup a channel with an open input for the desired ohms range and place the instrument
in monitor mode on that channel. Use a meter with high input impedance to measure the
open-circuit voltage at the channel input for the ohms range as listed in Table 5-6. If a
high input impedance meter is not available, only the 30-k
Ω
and lower ranges can be
checked.
Table 5-6. Ohms Open-Circuit Voltage
Range
Voltage
300
Ω
3 k
Ω
30 k
Ω
300 k
Ω
3 M
Ω
10 M
Ω
3V
1.3V
1.3V
3V
3V
3V
If the proper voltage is not measured, setup a channel on the 300
Ω
range (open input),
and have the instrument monitor that channel. Check for 3V dc with respect to A3TP9,
and work through the HI SOURCE and HI SENSE paths as described in Table 5-7.
If the HI path works correctly, trace continuity through the LO path. Check among A3L4
through A3L24, A3K1 through A3K14, A3R35, A3U8 pin 11, A3R43, A3K16, A3R34,
and A3U8 pin 13.
Table 5-7. Ohms HI Troubleshooting
Checkpoint
Signal Description
Possible Fault
A3U8 pin 14
Ohms Source
A3U8
A3R10 HI SRC
Ohms Source
A3R10, A3K16, A3RT1, A3Z4, A3Q10
Channel HI
Ohms Source
A3K1 through A3K14, A3U4, A3U5, A3U11, A3U12, A3L1,
A3L2, A3L3
A3U8 pin 23
Ohms Source
A3R11, A3K17, A3R42, A3C32
A3U8 pin 58
Ohms Source filter output
A3U8, A3Q2
Summary of Contents for 2620A
Page 4: ......
Page 18: ...HYDRA Service Manual 1 2...
Page 48: ...HYDRA Service Manual 1 32...
Page 140: ...HYDRA Service Manual 3 16...
Page 234: ...HYDRA Service Manual 5A 34...
Page 236: ...HYDRA Service Manual 6 2...
Page 254: ...HYDRA Service Manual 6 20 2620A 1601 s63f eps Figure 6 3 2620A 2625A A1 Main PCA...
Page 258: ...HYDRA Service Manual 6 24 2635A 1601 s64f eps Figure 6 4 2635A A1 Main PCA...
Page 260: ...HYDRA Service Manual 6 26 2620A 4002 CKT 1 CKT 2 s65f eps Figure 6 5 A2 Display PCA...
Page 272: ...HYDRA Service Manual 6 38...
Page 274: ...HYDRA Service Manual 7 2...
Page 282: ...HYDRA Service Manual 7 10...
Page 284: ...HYDRA Service Manual 8 2...
Page 285: ...Schematic Diagrams 8 8 3 Figure 8 1 A1 Main PCA 2620A 2625A 2620A 1601 s88f eps...
Page 288: ...Schematic Diagrams 8 8 6 Figure 8 1 A1 Main PCA 2620A 2625A cont 2620A 1001 3 of 4 s73f eps...
Page 292: ...Schematic Diagrams 8 8 10 2635A 1001 2 of 5 Figure 8 2 A1 Main PCA 2635A cont s76f eps...
Page 293: ...Schematic Diagrams 8 8 11 2635A 1001 3 of 5 Figure 8 2 A1 Main PCA 2635A cont s77f eps...
Page 296: ...Schematic Diagrams 8 8 14 2620A 4002 CKT 1 CKT 2 Figure 8 3 A2 Display PCA s90f eps...
Page 300: ...Schematic Diagrams 8 8 18 2620A 1003 2 of 3 Figure 8 4 A3 A D Converter PCA cont s82c eps...
Page 301: ...Schematic Diagrams 8 8 19 2620A 1003 3 of 3 Figure 8 4 A3 A D Converter PCA cont s83c eps...
Page 302: ...Schematic Diagrams 8 8 20 2620A 1604 Figure 8 5 A4 Analog Input PCA s92f eps...
Page 304: ...Schematic Diagrams 8 8 22 2620A 1605 Figure 8 6 A5 IEEE 488 Interface PCA 2620A Only s93f eps...
Page 306: ...Schematic Diagrams 8 8 24 2625A 1606 Figure 8 7 A6 Memory PCA 2625A s94f eps...
Page 310: ...Schematic Diagrams 8 8 28...