63
10 Digital Communication Interface
Interface Commands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Read CG calibration parameter
*cg
*cg
cg: 999.99
cg: 406.25
Set CG calibration parameter to
n
*cg=n
*cg=406.25
unlimited
Read Boost Heater mode
bo[ost]
bo
bo: xxxx
bo: auto
Set Boost Heater mode
bo[ost]=au[to]/us[er]
Auto/User
Set Boost to Auto mode
bo[ost]=au[to]
bo=au
Set Boost to User (manual) mode bo[ost]=us[er]
bo=us
Read low set-point limit value
*tl[ow]
*tl
tl: 999
tl: 0
Set low set-point limit to
n
*tl[ow]=n
*tl=0
0 to 20°C
Read high set-point limit value
*th[igh]
*th
th: 999
th: 205
Set high set-point limit to
n
*th[igh]=n
*th=205
30 to 300
Miscellaneous (not on menus)
Read all parameters
all
all
list of operating
parameters
Read firmware version number
*ver[sion]
*ver
ver.9999,9.99
ver.6331,1.03
Read all extended parameters
*all
*all
list of extended
parameters
Read structure of all commands
h[elp]
h
list of commands
Legend:
[] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note:
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Interface Command Summary continued