4.2
Set Up
WARNING:
The instrument is equipped with a soft cutout (user settable
firmware) and a hard cutout (set at the factory). Check the flash point,
boiling point, or other fluid characteristic applicable to the circumstances
of the instrument operation. Ensure that the soft cutout is adjusted to the
fluid characteristics of the application. As a guideline, the soft cutout
should be set 10°C to 15°C below the flash point of the bath fluid. See
Section 8.1, Heat Transfer Fluid, for specific information on bath fluids
and Section 9.10, Cutout.
Set up of the bath requires careful unpacking and placement of the bath, filling
the bath with fluid, and connecting power. Consult Section 5, Installation, for
detailed instructions for proper installation of the bath. Be sure to place the
bath in a safe, clean and level location. Refer to Section 5.3, Tipping Prevention
Bracket Installation.
Fill the bath tank with an appropriate fluid. For operation at moderate bath tem-
peratures, clean distilled water works well. Carefully pour the fluid into the
bath tank through the large rectangular access hole above the tank avoiding
spilling any fluid.
6331 Calibration Bath
User’s Guide
12