S
Set the new units and resume temperature display
9.6
Ramp and Soak Program
The ramp and soak program feature allows the user to program a number of
set-points and have the instrument automatically cycle between the tempera-
tures, holding at each for a length of time. The user can select one of four dif-
ferent cycle functions.
The program parameter menu is accessed by pressing “SET” and then “UP”.
| 100.00 C
Well temperature
S+U
Access program menu
ProG
Program menu
Press “SET” to enter the program menu
S
Enter program menu
9.6.1
Number of Program Set-points
The first parameter in the program menu is the number of set-points to cycle
through. Up to 8 set-points can be used in a ramp and soak program. These
set-points are independent from the programmable set-points described in Sec-
tion , Programmable Set-points.
Pn=8
Number of program set-points
Use the “UP” or “DOWN” buttons to change the number from 2 to 8.
Pn=3
New number of program set-points
Press “SET” to continue. Press “EXIT” to ignore any changes and to continue.
S
Save new setting
9.6.2
Set-points
The next parameters are the program set-points.
1 50.0
First set-point
43
9 Controller Operation
Ramp and Soak Program