FLIR
LEPTON® Engineering Datasheet
The information contained herein does not contain technology as defined by the EAR, 15 CFR 772, is publicly available,
and therefore, not subject to EAR. NSR (6/14/2018).
Information on this page is subject to change without notice.
Lepton Engineering Datasheet, Document Number: 500-0659-00-09 Rev: 203
80
Environmental stresses beyond those listed may cause permanent damage to the device. Exposure to absolute-
maximum-rated conditions for extended periods of time may affect device reliability.
Table 19 - Environmental Specifications
Stress
Maximum Rating
Operating Temperature Range
-10°C to 80°C
(-20°C to 80°C with some possible performance degradation)
Maximum Operating Temperature
80 °C1
Shutter Operating Temperature
-10°C to 80°C2
Storage Temperature
-40°C to 80°C
Altitude (pressure)
12 km altitude equivalent
Relative Humidity
95%
Thermal Shock
Air-to-air across operating temp. extremes (-10°C to 65°C,
65°C to -10°C)
Mechanical Shock
1500 g, 0.4 msec
Vibration
Transportation profile, 4.3 grms
ESD
Human Body Model (HBM), 2kV
Charged Device Model (CDM), 500V
Note(s)
1.
Lepton contains an automatic shutdown feature when its internal temperature exceeds the maximum safe operating value.
See
, page
2.
Lepton contains an automatic shutter lockout feature that prevents the shutter from operating when its internal temperature
is outside the range of -10°C to 80°C for Lepton 2.5 and 3.5, and -10°C to 65°C for Lepton 2.0 and 3.0. See
page
10.1
Compliance with Environmental Directives
Lepton complies with the following directives and regulations:
•
Directive 2002/95/EC,
“
Restriction of the use of certain Hazardous Substances in electrical and electronic
equipment (RoHS)”
•
Directive 2002/96/ EC,
“
Waste Electrical and Electronic Equipment (WEEE)
”
.
•
Regulation (EC) 1907/2006,
“
Registration, Evaluation, Authorization and Restriction of Chemicals
(REACH)”