Telemetry
Row
Word
start
Word
End
Number
of 16-bit
Words
Name
Notes
B
24
24
1
Window
Reflection
Scaled by 8192
B
25
25
1
Window
Temperature
Temperature in Kelvin x 100
B
26
26
1
Window
Reflected
Temperature
Temperature in Kelvin x 100
B
27
79
53
Reserved
C
0
4
5
Reserved
C
5
5
1
Gain Mode
6
0 = High, 1 = Low, 2 = Auto
C
6
6
1
Effective Gain
Mode
In Auto mode, 0 = High, 1 = Low
C
7
7
1
Gain Mode
Desired Flag
0 = current gain mode is desired, 1 = gain mode
switch desired
C
8
8
1
Temperature
Gain Mode
Threshold High to
Low (°C)
Temperature threshold in °C used to determine
when an Auto switch to Low gain mode (while in
High gain mode) should occur in Radiometry
enabled/TLinear disabled mode
C
9
9
1
Temperature
Gain Mode
Threshold Low to
High (°C)
Temperature threshold in °C used to determine
when an Auto switch to High gain mode (while in
Low gain mode) should occur in Radiometry
enabled/TLinear disabled mode
C
10
10
1
Temperature
Gain Mode
Threshold High to
Low (K)
Temperature threshold in Kelvin used to
determine when an Auto switch to Low gain mode
(while in High gain mode) should occur in TLinear
mode
6