Removal from service
For removing the Navigation System from service, shut down the system completely, and remove the main
power cord. Complete isolation of the Navigation Unit Cube
4D
is achieved only by disconnecting the power
connection from the power grid.
Cleaning and reconditioning
9.1. Cleaning the surface of the device
Manual cleaning is recommended. Turn off the Navigation Unit Cube
4D
at the main switch and remove the power
connection. Clean the whole surface of the Navigation Unit Cube
4D
with a dust-free soft cloth only if required.
The cloth may be moistened with water and a mild, soap-based, non-corrosive cleaning agent. Prevent liquids
from getting into the housing. For the disinfection of the surfaces, use an aldehyde-free quick-acting disinfectant
that allows wiping disinfection. Apply the solution onto a soft cloth and wipe the entire surface of the housing.
Wait until the solution has dried off before switching on the Navigation Unit Cube
4D
.
For cleaning the components and accessories (headrest, surgical instruments and the localizer), refer to the
cleaning recommendations included in the corresponding instructions for use.
9.2. Inspection Instructions
After cleaning, visually inspect the system under appropriate light conditions for remaining debris. Repeat
cleaning procedures if required. Refer to the Instructions For Use of the components and accessories for the
inspection recommendations related to the components and accessories.
CAUTION:
Never spray or pour cleaning agent or disinfectant directly onto the surface, because there
is a risk that the cleaning agent comes into direct contact with the components within the device and
causes irreparable damage to the electrical components. Spilling liquids on any electrical device
attached to the system should warrant inspection by an authorized Fiagon service representative.