28
Instructions for Use
Copyright © 2020
DocumentID
Version
Date
28 / 33
Fiagon GmbH
DOC-19-00169 2
08-2020
EN / US
12.3. Technical data
Navigation Unit Cube
4D
•
Energy consumption: 600 VA
•
Electric supply: 100 V-240 V AC, 50-60 Hz
•
Classification according to MDD 93/42/EEC: Class I
•
Protection class I, interface to applied part type BF
The device is not designed for use with or near inflammable or explosive gases.
The Fiagon Cube Navigation System WiFi works in the 2,4 GHz frequency band according to the standards
802.11 ac/a/b/g/n with a transmission power of 100 mW maximum.
Navigation sensor
•
Electromagnetic position measurement system
Operating
conditions
Temperature: +15°C to + 35°C,
Humidity: 30% to 75% without condensation
Air pressure: 700 hPa to 1060 hPa
Transport
conditions
Temperature: -10°C to + 35°C,
Humidity: 30% to 75% without condensation
Air pressure: 700 hPa to 1060 hPa
Storage
conditions
Protected against dust, and moisture.
Recommended storage temperature: +15° to +35°C
Recommended storage humidity: 30% to 75% without condensation
Recommended storage air pressure: 700 hPa to 1060 hPa
Warning:
If the Navigation System is transported indoors after outdoor placement, wait two hours
before use. Otherwise, condensed water on the components could compromise the proper functioning
of the system.
Essential Performance
•
Positioning accuracy on technical phantom (including registration process, mean value < 1,5 mm)
•
Distortion detection method. Field distortions are detected. (95% confidence interval is < 2mm)
•
Clinical total accuracy on anatomic preparation (mean value < 2 mm)