EMC compliance statement
The Fiagon Cube Navigation System needs special precautions regarding EMC and needs to be installed and
put into service according to the following described EMC information.
Electromagnetic emission
The Fiagon Cube Navigation System is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should assure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic Environment Guidance
RF-emissions following
IEC/CISPR 11
Group 1
The Fiagon Cube Navigation System uses RF energy
only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference
in nearby electronic equipment.
RF-emissions following
IEC/CISPR 11
Class B
The Fiagon Cube Navigation System is suitable for use
in all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Generation of
mains
harmonics following
IEC 61000-3-2
Class A
Generation of voltage
Fluctuations/ flicker following
IEC 61000-3-3
Complies