5.4. LED feedback
The Fiagon Cube Navigation System gives the user real-time feedback about its status and the status of
connected instruments and localizers via the LED feedback bar of the Navigation Unit Cube
4D
The following table states the status of different lighting:
LED bar signal
Status
Off – no emitting
System powered off
Fading in and out green
System is busy:
Software loading a dataset
Software calculating registration
Solid green
System is ready to be used:
•
Software startup completed
•
Software finished loading dataset
•
Software in Navigation Mode
Fading green-orange forth and back
System is waiting for user interaction:
•
Touch landmarks/Take photo/Hold to start surface
recording
•
Connect
mapper/camera
•
Confirm registration/use registration-ready instrument
Fading in and out orange
System warning:
•
Line feature is activated and the user gets close to the
set line
•
Localizer connected and distortion threshold reached or
localizer out of the electromagnetic field
Solid orange
System start:
•
System powered on and booting up
•
Software starting
•
Software in Mouse Mode
•
Field generator is connected and turned off/
•
Default state if software does not start or if the software
is closed
Solid red
Issue detected:
•
Broken instrument or localizer is detected in Mouse
Mode or Navigation Mode
•
Instrument with 0 uses remaining detected
The LED instrument ports give feedback about the connected instrument. The following table states the status
of different lighting:
LED instrument signal
Status
Off
Instrument disconnected or not recognized
Dim solid green
Instrument connected, recognized by the software, but non
active instrument
Solid green
Instrument connected, recognized by the software, and current
active instrument
Blinking red
Instrument/localizer connected, recognized by the software, but
broken or instrument has zero uses remaining
Solid blue
Patient localizer connected, recognized by software, and in
tracking field without distortion
Solid orange
Patient localizer detects distortion or out of tracking field.
Note: If the software application is not running and you connect an instrument or localizer, then the LED is still
off.