- 64 -
INF-TN1FSH-E
From condition 2
Total rate
≥
Full scale [m
3
/s]
×
2
×
Total pulse width [ms]
1000
= 0.01 [m
3
/s]
×
2
×
50 [ms]
1000
=
0.001 [m
3
] = 1 [L] ...................................B
The settable range of the total rate that satisfies both condition 1 and condition 2 is found to be as follows based on
the result of calculations A and B.
1 [L]
≤
Total rate
ii) In the case of DO3
From condition 1
Total rate
≥
Full scale [m
3
/s]
1 [Hz]
=
0.01 [m
3
/s]
1 [Hz] = 0.01 [m
3
] = 10 [L] ..........................................................C
Condition 2 is the same as the case of DO1 output in i) above.
Consequently, the settable range of the total rate is found to be as follows based on the result of calculations B and
C.
10 [L]
≤
Total rate
Operation
(example)
When starting integration with total rate set to 100 m
3
and pulse width 1 set to 100 msec.
Key operation
Description
Display
FUNC TOTAL
Display “TOTAL MODE”
TOTAL MODE
TOTAL STOP
#
Select “TOTAL RATE”
TOTAL RATE
0.000 m3
ENTER 1 0 0 ENTER
Enter numeric value enter mode, enter
“100” using ten keys, and press ENTER.
TOTAL RATE
100.000 m3
"
or
#
Select “PULSE WIDTH 1”
PULSE WIDTH 1
50 msec
ENTER
"
or
#
ENTER
Enter select/enter mode, select “100,” and
press ENTER.
PULSE WIDTH 1
100 msec
ESC
Return to “TOTAL MODE”
TOTAL MODE
TOTAL STOP
ENTER
"
or
#
ENTER
Enter select/enter mode, select “RESET,”
and press ENTER.
TOTAL MODE
TOTAL RESET
ESC
Display the measurement, reflecting the
setting.
(Measurement display screen)