Table 16 <Pulse and Doppler Measurement Detailed Setting>
Item Content
WEDGE SOUND
VELOCITY
With selection of AUTO/MANUAL, in case of MANUAL, input right side
column in the range of numeric 1000 to 3700 m/s.
PIPE SOUND VELOCITY With selection of AUTO/MANUAL, in case of MANUAL, input right side
column in the range of numeric 1000 to 3700 m/s.
LINING SOUND
VELOCITY
With selection of AUTO/MANUAL, in case of MANUAL, input right side
column in the range of numeric 1000 to 3700 m/s.
FLUID SOUND
VELOCITY
With selection of AUTO/MANUAL, in case of MANUAL, input right side
column in the range of numeric 500 to 2500 m/s.
TRANSMIT
FREQUENCY
With selection of AUTO/MANUAL, in case of MANUAL, input right side
column in the range of numeric 0.1 to 5 MHz.
SAMPLING
FREQUENCY
With selection of AUTO/MANUAL, in case of MANUAL, select numeric at right
side column.
RECEPTIVE WAIT TIME With selection of AUTO/MANUAL, in case of MANUAL, select numeric at right
side column.
REPETITION
FREQUENCY
With selection of AUTO/MANUAL, in case of MANUAL, input right side
column in the range of numeric 100 to 8000 Hz.
CHANNEL With
selection
of
AUTO/MANUAL,
in case of MANUAL, select from numeric
16, 32, 48, 64, 80, 96, 112 and 128 at right side column.
REFRENCE COUNT
Select numeric.
TRANSMIT PULSE NO.
Select from 0, 1, 2, 4, 8, 16, 32 and 64.
MEASUREMENT
RANGE
Select from F RADIUS, N RADIUS and DIAMETER.
PHASE ANGLE SHIFT
Select from NORMAL 1, NORMAL 2, POSITIVE and NEGATIVE.
GAIN
With selection of AUTO/MANUAL, in case of MANUAL, select numeric 0 to 18
in each column of START GAIN/END GAIN.
Table 17 <Pulse and Doppler Measurement Judgment Setting>
Item Content
SUCCESS RATIO
Enter in the range of 0 to 100%.
POWER
Enter in the range of 0.00 to 100.00.
DEVIATION
Enter in the range of 0.00 to 1.00.