FE FSH Instruction Manual Download Page 149

 

 

INF-TN1FSH-E

 

- 141 -

 

 

 DO2 

output 

confirmation 

(2)  Set [DO2 Output Confirmation] check box to ON.    Then select either ON or OFF from setting combo 

box, and click the [Setting] button to change to the selected value of DO2 output.    Click the [READ] 
button, and the setting value appears. 

 

 DO3 

output 

confirmation 

(3)  Set [DO3 Output Confirmation] check box to ON.    Then select either ON or OFF from setting combo 

box, and click the [Setting] button to change the selected value of DO3 output.  Click the [READ] 
button, and setting value is displayed. 

 

(3) Test 

mode 

Set input data and tracking time and click the [Setting] button, and you can enter the test mode. 
Click the [Read] button to read the values in each column of the test mode. 

 

[Release] button.............Resets analog output, each DO output and Test mode. 
* Note: Make sure to press the [Release] button when maintenance is completed. 

 
 

Summary of Contents for FSH

Page 1: ...Instruction Manual INF TN1FSHa E HYBRID ULTRASONIC FLOWMETER Duosonics TYPE FSH Flow transmitter FSW Detector FLY Signal cable...

Page 2: ...ouble caused by modification without permission If it becomes necessary to modify the flowmeter contact our office in advance This instruction manual should always be kept on hand by the operator Afte...

Page 3: ...product has not an explosion proof structure Do not use it in a place with explosive gases otherwise it can result in serious accidents such as explosion fire etc The unit should be installed in a pla...

Page 4: ...ed as far away as possible from high voltage lines to prevent entry of noise signals as it will cause malfunction of the unit To prevent malfunction of the unit the analog output signal cable and powe...

Page 5: ...A place not submerged 7 A place remote from electrical devices motor transformer etc which generate electromagnetic induction noise electrostatic noise etc 8 A place not subjected to excessive fluid...

Page 6: ...piping specifications 21 3 5 Installing Detector 23 3 5 1 Outline of detector installation procedure 23 3 5 2 Treatment of mounting surface 23 3 5 3 Mounting the detector by Z method using the frame 2...

Page 7: ...on 91 5 2 Periodic inspection 91 5 2 1 Checking zero point 91 5 2 2 Calibrating current output circuit 91 5 2 3 Calibrating temperature sensor circuit 91 5 2 4 Measuring insulation resistance 92 5 3 R...

Page 8: ...easurement 135 7 15 1 Detailed Setting optional function 135 7 15 2 Received Signal optional function 137 7 15 3 Operation Information 138 7 16 Maintenance 140 7 17 End 142 7 18 Uninstalling of Softwa...

Page 9: ...of clean liquids The new hybrid technique employing both the pulse Doppler and the transit time methods allows the flowmeter to be used for wider range of applications In addition our self developed...

Page 10: ...Z method Flow transmitter Power supply Temperature sensor cable 4 to 20 mA DC Contact 3 points max Detector Cable 2 2 measurement line method Z method Detector Cable Flow transmitter 4 to 20 mA DC Co...

Page 11: ...tor main unit FSWS50 1 set Absorber unit 1 set Wire rope 1 set Spring for mounting 1 set Signal cable FLY6 Cable for ultrasonic signals 1 pair 2 pcs Signal cable FLY7 Cable for temperature sensor 1 CD...

Page 12: ...for two path system Power supply 7th digit 100 to 240VAC 50 60Hz 20 to 30VDC Modification No 8th digit Case structure 9th digit Watertight type IP67 Conduit connection 10th digits G1 2 and G3 8 femal...

Page 13: ...Modification No 8th digit Mark 1 Signal cable 9th digit None Acoustic coupler 10th digit None Silicone compound Option 11th digit None Tag Plate Fuji Electric Systems Co Ltd Made in Japan Ser No Mfd T...

Page 14: ...e cable for temperature sonsor Cable length 5th to 7th digit 5m 10m 15m 20m 25m 30m 35m 40m 45m 50m 55m 60m 65m 70m 75m 80m 85m 90m 95m 100m 110m 120 130m 140m 150m Modification No 8th digit Mark1 6 7...

Page 15: ...or various setting operations 6 LCD Displays flow rate and various settings 7 Setting key Used for making settings 8 Power terminal Connect power cable to this terminal 9 I O terminal Connect power ou...

Page 16: ...less steel belt Used for fastening the sensor frame to the piping 19 Ultrasonic wave transmission surface Transmits ultrasonic waves to the fluid to be measured or receives them through the piping 20...

Page 17: ...2 A place not subject to direct sunlight or weather 3 A place provided with space for daily inspection and wiring work 4 A place not subject to radiant heat from a heating furnace etc 5 A place not in...

Page 18: ...measurement to a great extent Select a place that satisfies the conditions described in 2 2 1 Length of straight section of pipe Assure sufficient working space for installation and maintenance referr...

Page 19: ...ions shown below L 10D Detector L 50D L 30D In case that flow control valve exists on upstream side In case that flow control valve exists on downstream side L 10D L 10D L 10D L 5D L 30D 1 5D L 50D 0...

Page 20: ...with fluid The pipe may not be filled with fluid 2 Mount the flowmeter within 45 from the center plane in the case of horizontal pipe run Mount it at an arbitrary position on the outer periphery in th...

Page 21: ...sor is a reference The mounting spacing of the two sensors corresponds to the V shaped portion in Fig 3 Z method Retainer plate Sensor mounting spacing Flow direction Fig 1 Mounting of sensor Fig 2 Re...

Page 22: ...urement range is specified by parameter setting provided 8 Zero point adjustment Before performing zero point adjustment check that the pipe is filled with fluid the fluid is in still state and that t...

Page 23: ...an be mounted on a wall or on a panel Use four M8 bolts to mount the flow transmitter on a wall or a panel Drill holes according to the mounting hole dimensions shown below and fasten the flow transmi...

Page 24: ...sealing caps 3 3 2 Applicable wires Use the following cables Power cable 3 wire or 2 wire cabtyre cable Nominal sectional area 0 75 mm2 or more Finished outer diameter 11 mm Output signal cable 2 wir...

Page 25: ...2 M4 screws and remove the cover Remove the 4 M3 screws and then the shield plate On completion of wiring work be sure to mount the shield plate and the cover CAUTION Be sure to mount the cover before...

Page 26: ...9 2 3 4 5 7 7 7 7 7 2 2 4 2 2 4 2 2 4 2 3 4 3 4 Note 1 All screws of the terminal block are M3 5 Use solderless terminals with external diameter of 7 2 or less that are for M 3 5 Note 2 Both the termi...

Page 27: ...2 In sensor constant set the sensor constant calculated based on actual current calibration performed as part of the delivery test at the factory Set the sensor constant for each of the sensor units m...

Page 28: ...1 ENTER or ENTER Enter select enter mode select FSW12 and press ENTER SENSOR TYPE FSW12 Select SENSOR CONSTANT SENSOR CONSTANT SKIP ENTER or ENTER Enter select enter mode select Setting and press ENTE...

Page 29: ...l benzene bromide ethanol glycol kerosene milk methanol toluol lube oil fuel oil petrol and others Fluid S V 4 Numeric value 500 m s to 2500 m s Viscosity 4 Numeric value 0 0010E 6 m2 s to 999 9999E 6...

Page 30: ...25 using ten keys and press ENTER LINING THICKNESS 1 25 mm Select KIND OF FLUID KIND OF FLUID Water ENTER or ENTER Enter select enter mode select OTHERS and press ENTER KIND OF FLUID OTHERS Select FL...

Page 31: ...he piping to which the detector is to be mounted over the length of the frame to be used Note 1 If the pipe is wrapped with jute remove the jute wrapping over the entire circumference in width of fram...

Page 32: ...ting jig Note Pay attention when letting the shaft run through the jig so that the scale unit of the frame coincides with that of the shaft 6 Let the shaft run through the mounting jig on the opposite...

Page 33: ...f the stainless belt to let the tip of the belt run through it 11 Fasten the screw wrapping the stainless belt around the frame end 12 Fasten both ends of the frame with the stainless belt 13 Remove t...

Page 34: ...26 INF TN1FSH E 2 How to mount the frame not using a jig Gage paper is required to mount the frame by this method See 8 7 Making gauge paper for details 2 1 How to determine the mounting position 0 1...

Page 35: ...easurement lines Draw a mark off lines at an angle 45 from the horizon 2 Wrap the stainless belt around the pipe 3 Lift the screw of the stainless belt and let the tip of the belt run through it 4 Pla...

Page 36: ...ortion In this procedure fully fasten the screws of the retainer plate so that the sensor unit and absorber unit contact with the retainer plate as tightly as possible 2 Before mounting the sensor uni...

Page 37: ...nly one sensor to one frame 8 Connect the temperature sensor cable to the waterproof connector for the temperature sensor cable 9 Connect the ultrasonic signal cable to the BNC connector being careful...

Page 38: ...sten it using cloth belt etc not on the frame ends but on the frame 3 Then wrap a stainless steel belt around the frame end to fasten it securely Follow the description in 3 5 3 to fasten the stainles...

Page 39: ...or more Mount the sensor with wire in such cases Do not use a frame Perform the following to determine the mounting position Gauge paper is required for the work See 8 7 Making gauge paper for details...

Page 40: ...downstream sides and the absorber unit Allow the length of the wire rope to be the same or longer than the pipe diameter 3 Lay the wire rope around the pipe on the upstream side Then hang the spring f...

Page 41: ...ensor unit and the marking line The absorber unit is not required for resin piping Proceed to 11 in this case The absorber unit is used for metal piping Proceed to 8 in this case 8 Apply silicon rubbe...

Page 42: ...n the case of metal piping in the same manner 12 Connect the temperature sensor cable to the waterproof connector for the temperature sensor cable 13 Connect the ultrasonic signal cable to the BNC con...

Page 43: ...ect RANGE TYPE RANGE TYPE SINGLE ENTER or ENTER Enter select enter mode select Bi directional range and press ENTER RANGE TYPE BI DIR Select FULL SCALE 1 FULL SCALE 1 56 32 m 3 h ENTER 1 0 0 ENTER Ent...

Page 44: ...TE LIMIT 0 00 m3 h ENTER 5 ENTER Enter numeric value enter mode enter 5 and press ENTER RATE LIMIT 5 00 m3 h Select RATE LIMIT TIMER RATE LIMIT TIMER 0 sec ENTER 1 5 ENTER Enter numeric value enter mo...

Page 45: ...play screen Total setting example When performing integration holding output burnout with total pulse output set to 100 m3 for 1 pulse burnout timer set to 15 sec and pulse width 1 set as 100msec Key...

Page 46: ...ZERO Press Zero to perform zero adjustment in a state where the flow is stopped completely CLEAR Press Clear to perform zero adjustment in a state where the flow cannot be stopped Operation example Wh...

Page 47: ...or the upstream and the downstream of the sensor connection are reversed appears m3 s Display 1 Display 2 Measurement screen m s RAS display Measurement display Measurement method T For decimal point...

Page 48: ...ipe to which the sensor is to be mounted RANGE Output setting FUNC RANGE Used to set the analog output conditions unit range limit burnout rate limit DAMP Damping FUNC DAMP Used to set damping ZERO Ze...

Page 49: ...OTAL Total preset 4 4 9 3 Total switch 4 4 9 4 BURNOUT 4 4 9 5 Flow switch FUNC FLOW SW 4 4 10 Status output FUNC STATUS 4 4 11 Output calibration FUNC CAL 4 4 12 System unit 4 4 13 System language 4...

Page 50: ...ange unit 19 menus m s ft s m s L s L min L h L d kL d ML d m3 s m3 min m3 h m3 d km3 d Mm3 d BBL s BBL min BBL h BBL d KBBL d MBBL d ft s ft3 s ft3 min ft3 h ft3 d kft3 d Mft3 d gal s gal min gal h g...

Page 51: ...m s 0 00 to 104 98 ft s 4 00 m s 13 12 ft s 19 Unit 36 Flow switch Flow sw HYS 0 to 20 10 37 Output DO 1 15 menus Not used Not used Signal error F Total pulse R Total pulse F Total alarm R Total alarm...

Page 52: ...hod 2 Method 3 69 Signal balance 0 to 100 25 70 Transmission pattern 7 menus Burst 3 Burst 1 Burst 2 Burst 3 Burst 4 Burst 5 Chirp 4 Chirp 8 71 AGC gain 2 menus Auto Auto Manual 72 Signal peak 4 menus...

Page 53: ...be selected Selection Numeric selection Numeric values such as transmission voltage of the system can be selected Direct input example When entering outer diameter 114 3 mm Key operation Description...

Page 54: ...N STEEL PIPE MATERIAL CARBON STEEL ENTER Press ENTER PIPE MATERIAL CARBON STEEL Item selection example When changing trans voltage 80 Vpp to 160 Vpp Key operation Description Display Select TRANS VOLT...

Page 55: ...and FSW50 If hybrid method is selected sensor type can be selected only from FSW12 FSW21 FSW40 and FSW50 2 In sensor constant set the sensor constant calculated based on actual current calibration per...

Page 56: ...BRID Select SENSOR MOUNT SENSOR MOUNT V METHOD ENTER or ENTER Enter select enter mode select Z METHOD and press ENTER SENSOR MOUNT Z METHOD Select SENSOR TYPE SENSOR TYPE FSW21 ENTER or ENTER Enter se...

Page 57: ...bromide ethanol glycol kerosene milk methanol toluol lube oil fuel oil petrol and others Fluid S V 4 Numeric value 500 m s to 2500 m s Viscosity 4 Numeric value 0 0010E 6 m2 s to 999 9999E 6 m2 s 1 Se...

Page 58: ...IND OF FLUID WATER ENTER or ENTER Enter select enter mode select OTHERS and press ENTER KIND OF FLUID OTHERS Select FLUID S V FLUID S V 1440 m s ENTER 1 3 8 8 ENTER Enter numeric value enter mode ente...

Page 59: ...ocity 2 Line 1 velocity Line 2 velocity 2 Dumping Output correction Cut off Range Total Status output Rate limit Iout DO1 DO2 DO3 AO definition selector SW Flow rate calculation LCD 2 For 2 path mode...

Page 60: ...tional range d Bi directional auto 2 range 3 Setting range Set the full scale by range type Set the following depending on the range type The based scale is fixed to 0 Full scale Single range only Ful...

Page 61: ...eter FSW12 FSW21 FSW40 FSW50 FSW12 FSW21 FSW40 FSW50 40A 6 56 33 6 50A 6 52 52 7 65A 5 31 72 1 80A 4 65 86 5 90A 4 12 102 100A 3 69 7 25 118 231 125A 3 08 6 08 147 289 150A 2 63 5 20 179 354 200A 2 04...

Page 62: ...alue enter mode enter 100 using ten keys and press ENTER FULL SCALE 2 100 00 m 3 h Select HYSTERESIS HYSTERESIS 10 00 ENTER 5 ENTER Enter numeric value enter mode enter 5 using ten keys and press ENTE...

Page 63: ...upper limit 110 21 6 mA Key operation Description Display FUNC RANGE Display RANGE UNIT RANGE UNIT m 3 h or Select OUTPUT LIMIT Lo OUTPUT LIMIT Lo 20 ENTER 1 0 ENTER Enter numeric value enter mode ent...

Page 64: ...nout is not used Setting range of burnout timer 0 to 900 sec The burnout processing is performed as follows 1 LCD The measurement on the LCD changes with the analog output Operation example When outpu...

Page 65: ...it time is set to 0 this function does not work Operation example When rate limit is set to 5 m3 h and rate limit timer is set to 15 sec Key operation Description Display FUNC RANGE Display RANGE UNIT...

Page 66: ...ing range 0 0 to 100 0 63 Time Response time Flow rate Operation example When set value is 20 sec Key operation Description Display FUNC DAMP Display DAMPING DAMPING 5 0 sec ENTER 2 0 ENTER Enter nume...

Page 67: ...ample When zero adjustment is performed in a state where the flow is stopped Key operation Description Display FUNC ZERO Display ZERO ADJUSTMENT ZERO ADJUSTMENT CLEAR ENTER or Enter select enter mode...

Page 68: ...decimal places can be set arbitrarily within the display range Operation example When displaying instantaneous flow rate of display 1 in unit of m3 h and display 2 in unit of flow rate Key operation...

Page 69: ...off value as required Setting range 0 to 5 m s in velocity Enter absolute value Flow rate Cut off setting Output Operation example When the cut off point is set to 10 m3 h Key operation Description D...

Page 70: ...nter select enter mode select m3 and press ENTER TOTAL UNIT m3 ESC Return to TOTAL MODE TOTAL MODE TOTAL STOP ENTER or ENTER Enter select enter mode select TOTAL RESET and press ENTER TOTAL MODE TOTAL...

Page 71: ...msec 100 0 msec 200 0 msec DO3 Relay contact 1 pulse sec or lower 50msec 100 msec 200 msec The maximum output frequency is limited depending on the pulse width setting Set the total rate and the pulse...

Page 72: ...he result of calculations B and C 10 L Total rate Operation example When starting integration with total rate set to 100 m3 and pulse width 1 set to 100 msec Key operation Description Display FUNC TOT...

Page 73: ...cription Display FUNC TOTAL Display TOTAL MODE TOTAL MODE TOTAL STOP or Select F TOTAL PRESET F TOTAL PRESET 0 000 m3 ENTER 1 0 0 0 ENTER Enter numeric value enter mode enter 1000 using ten keys and p...

Page 74: ...mode is set to stop Total Set value ON OFF Status output Operation example When setting total switch setting in forward direction to 5000 m3 Key operation Description Display FUNC TOTAL Display TOTAL...

Page 75: ...is set to total stop 1 Integration is continued until output burnout processing is started Operation example When setting output burnout to hold and burnout timer to 30 sec Key operation Description D...

Page 76: ...the following range Set it in the range from 0 to 20 of the setting range Hysteresis is a ratio to the operation range Operation example When setting lower limit flow rate to 20 m3 h upper limit flow...

Page 77: ...ge is full scale 2 12 AO RANGE OVER Outputs when the value exceeds the upper limit setting or becomes lower than the lower limit setting of the range 13 PULSE RANGE OVER Outputs when the total pulse o...

Page 78: ...ENTER Select DO 2 and press ENTER SELECT STATUS DO 2 ENTER Display OUTPUT DO 2 OUTPUT DO 2 NOT USED ENTER or ENTER Enter select enter mode select F TOTAL PULSE and press ENTER OUTPUT DO 2 F TOTAL PUL...

Page 79: ...utput 0 Calibration of zero point Output 0 Calibration of span 100 Flow rate Flow rate Operation example When adjusting zero point to 0 5 m3 h and span to 105 Key operation Description Display FUNC CA...

Page 80: ...3 km3 Mm3 mBBL BBL kBBL Unit of viscosity E 6 m2 s Unit of temperature C Note Set units in a state where the total mode is set to stop Operation example When changing the unit system to metric system...

Page 81: ...en selecting English Key operation Description Display FUNC SYSTEM Display SYSTEM UNIT LANGUAGE SKIP ENTER or Select SETTING UNIT LANGUAGE SETTING ENTER Display SYSTEM UNIT SYSTEM UNIT METRIC or Displ...

Page 82: ...NC SYSTEM Display SYSTEM UNIT LANGUAGE SKIP or Select SERIAL COM SERIAL COM SKIP ENTER or Enter select enter mode and select SETTING SERIAL COM SETTING ENTER Display COM SPEED COM SPEED 38400 bps ENTE...

Page 83: ...KIP ENTER or ENTER Enter select enter mode select AO 1 and press ENTER AO 1 ADJUST 4 mA Display 4 mA adjustment Increase or Decrease Perform fine adjustment AO 1 ADJUST 4 mA Increase or Decrease Perfo...

Page 84: ...ation Description Display FUNC SYSTEM Display SYSTEM UNIT LANGUAGE SKIP or Select MAINTENANCE MAINTENANCE SKIP ENTER or ENTER Enter select enter mode select DO 1 and press ENTER DO 1 CHECK OFF ENTER o...

Page 85: ...le When performing calibration of resistance of wedge temperature measurement Key operation Description Display FUNC SYSTEM Display SYSTEM UNIT LANGUAGE SKIP or Select MAINTENANCE MAINTENANCE SKIP ENT...

Page 86: ...Select MAINTENANCE MAINTENANCE SKIP ENTER or ENTER Enter select enter mode select WEDGE TEMP and press ENTER ADJUST TEMP ADJUST or Select CHECK TEMP CHECK TEMP 0 0 C ENTER Update the temperature disp...

Page 87: ...elected as TOTAL MODE the total value also changes Select STOP not to make the total value change Operation example When setting target value to 100 and making it to be reached in 15 seconds Key opera...

Page 88: ...he backlight is set to ON when the change is made If the setting is changed from ON AUTO to OFF the backlight is set to OFF when the change is made Operation example When setting the backlight to AUTO...

Page 89: ...G and press ENTER SET PASSWORD 0000 ENTER 1 Enter numeric value enter mode enter 1 using ten keys and move to the second digit PASSWORD 1000 2 Enter 2 and move to the third digit PASSWORD 1200 3 Enter...

Page 90: ...me can be displayed Operation example Check the system name as follows Key operation Description Display FUNC SYSTEM Display SYSTEM UNIT LANGUAGE SKIP or Select SYSTEM NAME Check the system name SYSTE...

Page 91: ...g the description in 6 3 Checking received waveform CAUTION This parameter is intended for our service personnel Do not change the setting since the parameter affects the flow rate measurement If the...

Page 92: ...if the specified value is exceeded Factory set value 25 Numeric value 0 to 100 Note Set to 50 or higher for Method 1 Transmission pattern Setting of transmission pattern of ultrasonic signals Factory...

Page 93: ...NDOW CONTROL 2 WINDOW CONTROL AUTO ENTER or ENTER Enter select enter mode select MANUAL and press ENTER 2 WINDOW CONTROL MANUAL Select 2 OPEN TIME F 2 OPEN TIME F 0 us ENTER 1 5 0 ENTER Enter numeric...

Page 94: ...received waveform CAUTION This parameter is intended for our service personnel Do not change the setting since the parameter affects the flow rate measurement If the setting is changed measurement ma...

Page 95: ...taking in demodulated waves time from the start of transmission to the startup of window signals Numeric value Receipt wait time 0 12 s to 2104 75 s Repetition frequency Setting of frequency control m...

Page 96: ...ic value 0 00 to 99 99 E4 Deviation Setting of threshold value of standard deviation of Doppler shift Factory set value 0 5 A success rate error occurs if the threshold is exceeded The deviation measu...

Page 97: ...0 0 ENTER Enter numeric value enter mode enter 3500 using ten keys and press ENTER REPETITION FREQ 3500 Hz or Select LINE SELECT LINE SELECT LINE 1 F ENTER or ENTER Enter select enter mode select LINE...

Page 98: ...NC DETAIL Display VERSION INF VERSION INF SKIP or Select SETTING DATA SETTING DATA NOT INITIALIZE ENTER or ENTER Enter select enter mode and then select FACTORY SETTING and the converter is reset SETT...

Page 99: ...l Check that the pipe is filled with fluid Decrease air bubbles or foreign substances if contained in the fluid too much 5 2 Periodic inspection 5 2 1 Checking zero point Stop the flow of the fluid fi...

Page 100: ...terminal and grounding terminal In the case of AC power Between L N batch and outer earth terminal In the case of DC power Between batch and outer earth terminal 2 Between output terminal and groundin...

Page 101: ...3A such as UL CSA FGMT 250 V 3 A by Fuji Tanshi Kougyo 1 Turn off the power and open the cover Loosen the 4 screws on the front face of the flow transmitter and open the cover 2 Replacement of fuse Re...

Page 102: ...shown by the following photo pull out the card relay on the measurement board located under the power supply board from the socket 3 Set a new card relay into the socket Push the card relay securely...

Page 103: ...or 4 Remove the screws fastening the LCD unit 4 positions 5 Mount a new LCD unit see parts list Insert the operation key into the hole of the cover properly paying attention not to let the operation k...

Page 104: ...positive direction Flow in negative direction Displayed symbol Operation status None Normal E1 Device error backup memory wedge temperature measurement circuit E2 Measurement error transit time error...

Page 105: ...und waves cannot be propagated into pipe E2 Signal error Check receive sensitivity Check mixing in of air bubbles Check mixing in of foreign substances Check zero point status Receive sensitivity is l...

Page 106: ...00010000000000000 E2 NO SIGNAL or Select multiple error items if any by pressing or ENTER Check the contents of the error selected If multiple error items are selected is displayed RAS INFORMATION E2...

Page 107: ...N INF SKIP or Select RAS RAS SKIP ENTER or ENTER Set to Select input mode select RAS SETTING and press ENTER RAS SETTING 11 111111 1111 11 E1 DEVICE ERROR ENTER Set the selected anomaly to 0 Valid 1 I...

Page 108: ...CTION RESERVE DEVICE ERROR RESERVE RESERVE RESERVE RESERVE RESERVE PULSE RANGE OVER AO RANGE OVER F TOTAL SW R TOTAL PULSE Operation example In case of range over Key operation Description Display FUN...

Page 109: ...Wedge temperature C Wedge temperature C Angle in fluid Theoretical value Angle in fluid Theoretical value Total time s Theoretical value Transmission frequency MHz Window open s Theoretical value Sam...

Page 110: ...LINE SELECT NO LINE 1 ENTER or ENTER Enter select enter mode select LINE 2 and press ENTER 2 TOTAL TIME 89 256 us or Select SIGNAL POWER R and check the data 2 SIGNAL POWER R 56 23 ESC ESC Go back to...

Page 111: ...the grounding terminal on the flow transmitter case To 6 3 4 Measures against hardware failure Pale display Ambient temperature is low less than 20 C When temperature cannot be increased The LCD has c...

Page 112: ...silicon at the time of sensor mounting Improper connection of sensor cable Check and remove the sensor apply silicon filling material again and mount the sensor in a position slightly deviated from t...

Page 113: ...formed Perform zero adjustment in a state the flow is completely stopped The reading does not appear as 0 even if the flow is stopped If the flow is stopped the fluid does not fill the pipe or the pip...

Page 114: ...s 20 mA OVER FLOW appears on the LCD Range over Set the analog output range data once again The output becomes lower than 4 mA UNDER FLOW appears on the LCD Reverse flow Set upper lower stream properl...

Page 115: ...nel for the work described below 6 3 1 Method by oscilloscope Open the cover and connect an oscilloscope to the check pin on the printed board according to the following figure The unit has high volta...

Page 116: ...When it is larger than 2 4 Vpp The flow transmitter may be defective Inform us of the details of the error 3 Check that overall noise level is kept at 0 48 Vpp or lower If the noise level is higher t...

Page 117: ...y on LCD of 6 2 3 Measurement value error 2 The peak amplitude does not fluctuate If the peak fluctuates vertically air bubbles may be mixed in Take measures by referring to the part of mixing in of a...

Page 118: ...Echo sensitivity is low See the following figure 2 When amplitude is larger than 3 2 Vpp The flow transmitter may be defective Inform us of the details of the error When there are no Doppler signals...

Page 119: ...data can be easily edited with Microsoft Excel because you can save your data in CSV file format Note Microsoft Excel is the registered Trademark of the Microsoft Corporation in the United States 7 3...

Page 120: ...ancel button to cancel the installation Fig 5 Setup wizard screen 3 There is a query about selection of installation folder Click the Next button to install the software in that folder To specify a fo...

Page 121: ...ation Click the Next button to execute the installation Click the Previous button to return to the previous screen Click the Cancel button to cancel the installation Fig 7 Installation confirmation sc...

Page 122: ...omplete screen is displayed Click the Close button to exit the installation screen Fig 9 Installation complete screen 7 After installation the start menu and the application Duosonics_ENG that has bee...

Page 123: ...ge and unit can be obtained by communicating with the flow transmitter If error occurs during communications an error message is displayed to continue communication select Continue To stop communicati...

Page 124: ...unit is obtained Click the Cancel button to invalidate the setting Table 2 Measurement and Detailed Setting Item Content Port No Select either from COM1 COM2 COM3 COM4 and COM5 Serial Method Select e...

Page 125: ...tton Click the Cancel button not to save the setting File format is ini file Fig 13 Save setting select save file screen Note Please be careful not to rewrite the setting file for loader Hybrid USF in...

Page 126: ...range related matters TOTAL Sets total related matters STATUS Sets status output related matters DISPLAY Sets LCD display related matters SYSTEM Sets system related to language etc MEASURE Displays tr...

Page 127: ...d as fluid type fluid sound velocity and dynamic viscous coefficient become valid Setting Sends the setting of the selected item check box set to ON reflecting the response value on the setting READ R...

Page 128: ...VISCOSITY Enter in the range from 0 0010 to 999 9999 10 6 m2 s if fluid type is Others SENSOR SPACING Read only is valid Table 4 System Item Content MEASURE METHOD Select from hybrid and transit time...

Page 129: ...rse range forward and reverse automatic 2 range Display Valid Full scale 1 full scale 2 and histeresis Display Invalid Full scale Setting Sends the setting of the selected item check box set to ON ref...

Page 130: ...TPUT LIMIT LOW Enter in the range of 20 to 0 OUTPUT LIMIT HIGH Enter in the range of 100 to 120 OUTPUT BURNOUT Select from NOT USED HOLD UPPER LOWER ZERO BURNOUT TIMER Enter in the range of 0 to 900se...

Page 131: ...eflecting the response value on the setting Note that only when STOP mode is selected the setting of other items is reflected READ Reads the setting of the selected item check box set to ON reflecting...

Page 132: ...o 9999999999 999 R TOTAL PRESET Enter in the range of 0 to 9999999999 999 R TOTAL SW Enter in the range of 0 to 9999999999 999 OUTPUT BURNOUT Select from NOT USED and HOLD BURNOUT TIMER Enter in the r...

Page 133: ...ting Save Reflects the setting sent by pressing the Setting button on the flowmeter flow transmitter Check ON OFF Set the check box to ON to select all the items to set all the check boxes to ON Set t...

Page 134: ...setting sent by pressing the Setting button on the flowmeter flow transmitter Check ON OFF Set the check box to ON to select all the items to set all the check boxes to ON Set the check box to OFF to...

Page 135: ...cting the response value on the setting READ Reads the setting of the selected item check box set to ON reflecting the response value on the setting Save Reflects the setting sent by pressing the Sett...

Page 136: ...ldest is deleted and time is shifted to make the latest value to be seen when specified points are reached The vertical axial it displays with Y scale specifed The grid line represents Pulse Doppler i...

Page 137: ...tting screen To select an item to be set or read set the relevant check box to ON Not to select or to reset the selection set the relevant check box to OFF As to judgment it can obtain with setting su...

Page 138: ...selection of AUTO MANUAL in case of MANUAL select numeric at right side column RECEPTIVE WAIT TIME With selection of AUTO MANUAL in case of MANUAL select numeric at right side column REPETITION FREQU...

Page 139: ...bution measured by down flow side sensor in red Collection Moment Displays data by each read Moving Average Displays data in moving average with the number of times set by channel in each read data Mo...

Page 140: ...132 INF TN1FSH E Fig 25 Flow Velocity Distribution screen Internal pipe wall of upper stream sensor Center of piping Internal pipe wall of down stream sensor...

Page 141: ...ement method Fig 26 Operation Information screen Select either Line 1 or Line 2 first READ Reads operation information in batch Save As CSV Saves the measurement result in a CSV format file Click the...

Page 142: ...normal ANGLE IN FLUID TRANSMIT FREQUENCY MHz SAMPLING FREQUENCY kHz RECEPTIVE WAIT TIME s REPETITION FREQUENCY Hz TRANSMIT PULSE No REFERENCE COUNT No OF CHANNELS MEASURE RANGE F RADIUS N RADIUS DIAME...

Page 143: ...NG and the following screen appears Fig 27 Detailed information screen Select either Line 1 or Line 2 first As to selected Lateral line select the items to be set and read Set the check box items to b...

Page 144: ...in the range of numeric 0 to 100 SIGNAL PEEK Select from 2048 3071 4096 and 5120 TRIGGER LEVEL With selection of AUTO MANUAL in case of MANUAL input range of numeric 10 00 to 90 00 at right column WI...

Page 145: ...e and the selected range is magnified Press the R key to return to original status Method 1 One from forward direction reverse direction and correlation waveform can be selected Method 2 One from forw...

Page 146: ...en Select either Line 1 or Lline 2 first Read Reads operation information in a batch Save As CSV Saves Operation Information in file with CSV format Click the button and you are prompted to enter the...

Page 147: ...E IN LINING FLUID SOUND VELOCITY m s ft s WEDGE TEMPERATURE C F displaying with in case of measurement abnormal ANGLE IN FLUID TOTAL TIME s WINDOW OPEN s TOTAL TIME s FORWARD TIME s RESERVE TIME s TRA...

Page 148: ...olumn and click the Setting button and then click the Save button When setting is completed setting value is redisplayed at right column Click the READ button and selected setting values of 4 mA and 2...

Page 149: ...tion check box to ON Then select either ON or OFF from setting combo box and click the Setting button to change the selected value of DO3 output Click the READ button and setting value is displayed 3...

Page 150: ...tting appears To save setting value select Yes On the file designation window that appears select a file and the setting is saved in the file Then the loader is terminated Not to save setting value se...

Page 151: ...Stop bit 1 2 bits BCC Even horizontal parity Station 01 to 31 Number of connected units 31 max system including other devices 8 1 2 Message configuration 8 1 2 1 Receiving Configuration Byte count No...

Page 152: ...onse format STX SLV U P F_CD Data BCC CR LF Error response format STX SLV U P F_CD ERR BCC CR LF 8 1 3 Error check EX OR even horizontal parity from STX side of each byte excluding STX BCC CR and LF O...

Page 153: ...H 20 4 Status information H 20 5 Total value Forward direction 0004 1 Forward flow rate integration MDV 15 Number of decimal places 3 Total unit 2 Measurement method H 2 Time difference 1 Pulse Doppl...

Page 154: ...Format error FORMAT ERROR 12 characters MANAGE ERROR 12 Management error MANAGE ERROR 12 characters REQUEST ERROR 13 Request error REQUEST ERROR 13 characters 8 1 6 Cable connection specifications RS...

Page 155: ...maximum measurement range in Hybrid mode refer to the next page Transit time method Pipe size inside diameter Velocity Accuracy 2 m s to 32 m s 1 5 of rate 40mm to 50mm or less 0 m s to 2 m s 0 03m s...

Page 156: ...s Operation on PC98 series machines NEC cannot be guaranteed Operation on self made PCs or shop brand PCs cannot be guaranteed Major functions Setting changing of various parameters for the main unit...

Page 157: ...INF TN1FSH E 149 8 3 Outline diagram Flow transmitter Type FSH With waterproof gland 134 3 Mtg plate 4 9 Duosonics 352 332 240 72 233 14 33 33 30 30 30 24 48 48 48...

Page 158: ...150 INF TN1FSH E With union gland 134 3 Mtg plate 4 9 Duosonics 352 332 240 72 233 14 33 33 30 30 30 24 48 48 48...

Page 159: ...Sensor unit without temp sensor Cursor Absorber unit Frame FSWS12 TYPE FSWS21 40 to 200 360 PIPE SIZE L 100 to 400 540 MASS APPROX 1 7 57 70 W H 57 72 1 9 Detecter Type FSWS40 Temp sensor BNC connect...

Page 160: ...152 INF TN1FSH E Detecter Type FSWS50 Temp sensor BNC Connector Sensor unit with temp sensor Sensor unit without temp sensor Absorber unit 258 73 82...

Page 161: ...ing to the designation of the 5th 6th and the 7th digits of the Code Symbols 21 5 6 9 Connector 3 6 6 L 1 m Loader cable ZZP TK4H6253 RS 232C Terminal cap Red Terminal cap Yellow To Flow transmitter T...

Page 162: ...2 s 0 0107 to 10763 9088 10 6 ft2 s Water Water Seawater DIST water Ammonia Alcohol Benzene Bromide Ethanol Glycol Kerosene Milk Methanol Toluol Lube oil Fuel oil Petrol Others Sound velocity m s ft s...

Page 163: ...Total Pulse width 2 9 menus 50 0 ms 0 5 1 0 2 0 5 0 10 0 20 0 50 0 100 0 200 0 33 Flow sw low In terms of flow velocity 0 00 to 32 00 m s 0 00 to 104 98 ft s 0 00 m s 0 00 ft s 19 Unit 34 Flow sw high...

Page 164: ...the detector specification 1 Path 2 Path 50 Measurement mode AO definition 3 menus Depends on the detector specification Average Line 1 Line 2 51 System Sensor Type 4 menus FSW12 FSW12 FSW21 FSW40 FS...

Page 165: ...WALL THICKNESS LINING MATERIAL NO LINING TAR EPOXY MORTAR RUBBER TEFLON PYREX GLASS PCV OTHERS LINING S V Note Displayed when OTHERS is selected from the lining materials LINING T Note Displayed when...

Page 166: ...LIMIT LOW OUTPUT LIMIT HIGH BURNOUT NOT USED HOLD UPPER LOWER ZERO BURNOUT TIMER Note Displayed when those other than NOT USED are selected RATE LIMIT RATE LIMIT TIMER DAMPING FUNC DAMP DAMPING ZERO A...

Page 167: ...1 0 2 0 5 0 10 0 20 0 50 0 100 0 200 0 FLOW SWITCH FUNC FLOW SW FLOW SW LOW FLOW SW HIGH HYSTERESIS STATUS OUTPUT FUNC STATUS SELECT STATUS DO 1 OUTPUT DO 1 NOT USED SIGNAL ERROR F TOTAL PULSE R TOTA...

Page 168: ...D V METHOD MEASUREMENT MODE 1 PATH 2 PATH AO DEFINITION LINE 1 LINE 2 Note Displayed when 2 PATH is selected for MEASUREMENT MODE AVERAGE Note Displayed when 2 PATH is selected for MEASUREMENT MODE SE...

Page 169: ...s LINE 1 Note Represents LINE No PULSE DOPPLER SKIP CHECK WEDGE S V WEDGE ANGLE PIPE S V ANGLE IN PIPE LINING S V ANGLE IN LINING FLUID S V WEDGE TEMP ANGLE IN FLUID TRANS FREQUENCY SAMPLING FREQ RECE...

Page 170: ...E R SATURATION MEAS METHOD METHOD 1 METHOD 2 METHOD 3 SIGNAL BALANCE TRANS PATTERN BURST 1 BURST 2 BURST 3 BURST 4 BURST 5 CHIRP 4 CHIRP 8 RESERVE AGC GAIN AUTO MANUAL AGC LEVEL F AGC LEVEL R SIGNAL P...

Page 171: ...RADIUS DIAMETER PHASE ANGLE SHIFT NORMAL 1 NORMAL 2 POSITIVE NEGATIVE GAIN AUTO MANUAL START GAIN END GAIN LINE SELECTION LINE 1 F POWER DEVIATION SUCCESS RATE LINE 1 R Same as LINE 1 F LINE 2 F Same...

Page 172: ...2 23 0 250 10 267 4 3 4 4 0 6 5 9 3 15 1 21 4 28 6 300 12 318 5 4 0 4 5 6 5 10 3 17 4 25 4 33 3 350 14 355 6 11 1 19 0 27 8 35 7 400 16 406 4 12 7 21 4 30 9 40 5 450 18 457 2 14 3 23 8 34 9 45 2 500...

Page 173: ...52 700 49 798 39 778 30 760 800 56 912 44 888 34 868 900 49 998 38 976 1000 54 1108 42 1084 1100 59 1218 46 1192 1200 65 1330 50 1300 1300 73 1496 57 1464 1500 81 1662 63 1626 Polyethylene pipe for ge...

Page 174: ...16 10 3 216 6 5 250 267 12 7 267 7 8 300 318 15 1 318 9 2 350 370 10 5 400 420 11 8 450 470 13 2 500 520 14 6 600 630 17 8 700 732 21 0 800 835 23 9 Carbon steel pipe for pipe arrangement JIS G3452 19...

Page 175: ...0 15 0 18 0 15 0 2100 2133 6 19 0 16 0 19 0 16 0 2200 2235 2 20 0 16 0 20 0 16 0 2300 2336 8 21 0 17 0 21 0 17 0 2400 2438 4 22 0 18 0 22 0 18 0 2500 2540 0 23 0 18 0 23 0 18 0 2600 2641 6 24 0 19 0...

Page 176: ...12 7 17 5 900 36 914 1 8 0 12 7 19 1 1000 40 1016 0 9 5 14 3 26 2 Ductile iron specials Nominal diameter mm Thickness mm 75 8 5 100 8 5 150 9 0 200 11 0 250 12 0 300 12 5 350 13 0 400 14 0 450 14 5 5...

Page 177: ...5 187 650 26 660 4 96 8 103 114 127 140 152 165 178 190 203 700 28 711 2 104 111 123 137 151 164 178 192 205 219 750 30 762 0 119 132 147 162 176 191 206 220 235 800 32 812 8 127 141 157 173 188 204 2...

Page 178: ...59 35 1520 12 60 1551 30 85 1553 25 11 1451 51 36 1522 06 61 1551 88 86 1552 82 12 1455 34 37 1523 93 62 1552 42 87 1552 37 13 1459 07 38 1525 74 63 1552 91 88 1551 88 14 1462 70 39 1527 49 64 1553 35...

Page 179: ...temperature density V sound velocity c Sound velocity by piping material Material V m s Iron 3230 Steel 3206 Ductile cast iron 3000 Cast iron 2460 Stainless steel 3206 Copper 2260 Lead 2170 Aluminum...

Page 180: ...aving the length of 4D and width of 200 mm D if possible or longer with long sides parallel to each other 4D or more 200 mm or D or more D Pipe diameter Parallel 2 Draw a line that intersects with the...

Page 181: ...ast Tower 11 2 Osaki 1 chome Shinagawa ku Tokyo 141 0032 Japan http www fesys co jp eng http www fic net jp eng Phone 81 42 585 6201 6202 Fax 81 42 585 6187 Sales Div III International Sales Group Glo...

Reviews: