INF-TN1FSH-E
- 59 -
4.4.6. Zero adjustment
Description
The zero point of the measured value by time difference measurement can be adjusted as follows.
(Setting contents)
Zero: Perform zero adjustment in a state where the flow is stopped.
The measurement status at the specified time is set as 0.
Note: Perform adjustment in a state where the flow is stopped.
Note: Perform adjustment in normal measurement status.
Clear: Use Clear when the flow cannot be stopped.
Clears the value “adjusted.”
Operation
(example)
When zero adjustment is performed in a state where the flow is stopped.
Key operation
Description
Display
FUNC ZERO
Display ZERO ADJUSTMENT.
ZERO ADJUSTMENT
CLEAR
ENTER
"
or
#
Enter select/enter mode, and select “ZERO”
ZERO ADJUSTMENT
ZERO
▼
▲
ENTER
Adjustment is performed.
Elapsed time is displayed in the lower row while
adjustment is in progress.
ZERO ADJUSTMENT
######
When adjustment is completed successfully, “ZERO”
is displayed, and when it is completed unsuccessfully,
“CLEAR” is displayed on the lower row.
ZERO ADJUSTMENT
ZERO
ESC
Display the measurement, reflecting the setting.
(Measurement display screen)
CAUTION
•
If “CLEAR” is selected and executed, currently stored zero adjustment value is cleared.