4000 Series
Page
25
of
41
3460006-17_serial series 5000000
Seconds
0
Rate
0ºC
Wait for Process Instance
1
Must always be 1 for this step type
Wait For Process 1
-80ºC
The temperature waiting to be reached
Wait Event 1
None
Wait Event 2
None
Day of Week
Every Day
Jump Step
1
Jump Count
1
End Type
User
Event 1
On
Tied to internal outputs on the controller
Event 2
Off
Event 1 must be ON and Event 2 OFF
for this step type
Wait for Event
This step type is configured in a way that will hold the previous steps set point until a user presses the CYCLE
STOP button. This step will only be used at the end of a profile.
Step Type
Wait for Event
Target Set Point Loop 1
40ºC
Target Set Point Loop 2
-18ºC
Hours
0
Minutes
0
Seconds
0
Rate
0ºC
Wait for Process Instance
1
Wait For Process 1
40ºC
Wait Event 1
On
Wait Event 2
Off
Day of Week
Every Day
Jump Step
1
Jump Count
1
End Type
User
Event 1
On
Event 2
On
End
An end step will turn off all functions in the unit, and return the unit to the standby state. The temperature in the unit
will no longer be controlled while in the stand by state.
Step Type
End
Target Set Point Loop 1
-18ºC
Target Set Point Loop 2
-18ºC