4000 Series
Page
27
of
41
3460006-17_serial series 5000000
Profile Examples
Default profile 1 example
Click on Profile1 step then Profile 1 Step 1. This step is the first to use when configuring profile 1. Note,
the only options available for the Step Type (Time) are Target Set Point 1, Hours, Minutes, Seconds,
Event 1 and Event 2. All others are grayed out. When selecting other Step Types other options
selectable will be in dark characters.
Screen 5
Screen 5 description: Profile 1 Step 1 is a time step and will ramp down to the -80C set point in 10
seconds. Event 1 is on, turning on the refrigeration liquid, hot gas solenoid valves and turning on the profile
running indicator on the mode switch. Event 2 is off.
Screen 6