3460006-17_serial series 5000000
Screen 6 description: Profile 1 Step 2 is a wait for process step, the Wait for Process Instance 1
monitors the chamber temperature to reach the Wait for Process 1 temperature of -80C. Event 1 is on
which continues to activate the refrigeration liquid, hot gas solenoid valves and profile running indicator
on the mode switch. Event 2 is off. Note the only selectable options for this step are Wait for Process
Instance, Wait for Process 1, Event 1 and Event 2.