FaroArm
®
Quantum
August 2017
65
Chapter 3: Operation
CAUTION
: The probe
must
be well-seated in the hole when digitizing all
compensation points. Even one or two poorly digitized points significantly
affects the optimization process, which then has an effect on the accuracy of
the Quantum.
6.
Digitize points in the compensation
cone and sweep to position #3.
•
Press and hold the green
Front
button and move the handle down
until the shaft of the probe is
parallel with the top of the cone.
N
OTE
: You must digitize at least 150
points.
N
OTE
: Seven-axis Quantums have an additional position. Rotate the
handle.
7.
Digitize points in the compensation
cone and rotate the handle.
•
Press and hold the green
Front
button and rotate the handle.
N
OTE
: You must digitize at least 50
points in this position.
8.
Press the red
Back
button to finish.
08m80e00_FaroArm_Quantum.book Page 65 Tuesday, August 22, 2017 11:03 AM
Summary of Contents for FaroArm Quantum
Page 2: ...08m80e00_FaroArm_Quantum book Page 2 Tuesday August 22 2017 11 03 AM...
Page 4: ...08m80e00_FaroArm_Quantum book Page 2 Tuesday August 22 2017 11 03 AM...
Page 10: ...08m80e00_FaroArm_Quantum book Page vi Tuesday August 22 2017 11 03 AM...
Page 128: ...08m80e00_FaroArm_Quantum book Page 118 Tuesday August 22 2017 11 03 AM...
Page 154: ...08m80e00_FaroArm_Quantum book Page 144 Tuesday August 22 2017 11 03 AM...
Page 170: ...08m80e00_FaroArm_Quantum book Page 160 Tuesday August 22 2017 11 03 AM...
Page 174: ...164 08m80e00_FaroArm_Quantum book Page 164 Tuesday August 22 2017 11 03 AM...
Page 190: ...08m80e00_FaroArm_Quantum book Page 6 Tuesday August 22 2017 11 03 AM...
Page 200: ...08m80e00_FaroArm_Quantum book Page 6 Tuesday August 22 2017 11 03 AM...
Page 201: ...08m80e00_FaroArm_Quantum book Page 1 Tuesday August 22 2017 11 03 AM...