FaroArm
®
Quantum
August 2017
100
Chapter 4: Probes
2.
Click
Start Compensation
. Digitize the white area of the FARO
Compensation Plate with the FARO Laser Line Probe in a sweeping motion.
You should collect at least 200 scan lines in the FARO Laser Line Probe
compensation. If not, repeat the compensation moving the FARO Laser Line
Probe slower in each step.
•
Aim the laser line at the middle of the white surface.
•
Move the FARO Laser Line Probe until the laser is in the
Center Range
.
•
Press and hold the green
Front
button and move the FARO Laser Line
Probe while pointing to the center of the white surface. Rotate the
FARO Laser Line Probe as much as possible - at least 90 degrees.
•
Press the red
Back
button.
Figure 4-15 Measuring the Compensation Plate with the Laser Line Probe
08m80e00_FaroArm_Quantum.book Page 100 Tuesday, August 22, 2017 11:03 AM
Summary of Contents for FaroArm Quantum
Page 2: ...08m80e00_FaroArm_Quantum book Page 2 Tuesday August 22 2017 11 03 AM...
Page 4: ...08m80e00_FaroArm_Quantum book Page 2 Tuesday August 22 2017 11 03 AM...
Page 10: ...08m80e00_FaroArm_Quantum book Page vi Tuesday August 22 2017 11 03 AM...
Page 128: ...08m80e00_FaroArm_Quantum book Page 118 Tuesday August 22 2017 11 03 AM...
Page 154: ...08m80e00_FaroArm_Quantum book Page 144 Tuesday August 22 2017 11 03 AM...
Page 170: ...08m80e00_FaroArm_Quantum book Page 160 Tuesday August 22 2017 11 03 AM...
Page 174: ...164 08m80e00_FaroArm_Quantum book Page 164 Tuesday August 22 2017 11 03 AM...
Page 190: ...08m80e00_FaroArm_Quantum book Page 6 Tuesday August 22 2017 11 03 AM...
Page 200: ...08m80e00_FaroArm_Quantum book Page 6 Tuesday August 22 2017 11 03 AM...
Page 201: ...08m80e00_FaroArm_Quantum book Page 1 Tuesday August 22 2017 11 03 AM...