4) Press SET key.
Illuminating part switches from the measured value
display lamp to the set value display lamp and the
current set temperature is displayed on the inlet
temperature display.
5) Change the desired inlet temperature with UP key
and DOWN key.
6) Press the SET key to confirm the set value.
Illuminating part switched from the set value display
lamp to the measured value display lamp and the
current measured value (inlet temperature) is
displayed on the inlet temperature display.
*The setting range of inlet temperature is 40~200ºC.
Operate the unit within the setting range.
5) Press and hold Temp. Controller/Blower RUN/STOP
key (1 sec).
The temperature controller and blower start running.
*Temp. Controller/Blower operation lamp illuminates
during operation of temperature controller/blower.
*The heating lamp illuminates during heating (heater
turned ON)
*If the inlet temperature is heated beyond its set
temperature and the set tempe>10ºC
continues for 3 min, the abnormal temperature
alarm works and all operations excluding agitation
stop.
(The heater alarm lamp illuminates)
(Buzzer keeps ringing for 30 sec)
*If during heating the current flown into the heater
lowers below the standard, the abnormal heater
alarm works and all operations excluding agitation
stop.
(The heater alarm lamp illuminates)
(Buzzer keeps ringing for 30 sec)
*If dry air volume is insufficient, the overheating
protector may start up. During actuation, the
overheating protector actuation lamp illuminates in
the back of the left side of the unit.
To deactivate the overheating protector, increase the
dry air volume with the dry air volume adjusting
knob to lower the temperature at the heater and
press the overheating protector actuation RESET
switch. Once the heater is fully cooled down, the
overheating protector actuation lamp goes out.
Outlet temp.indicator
Inlet temp. indicator
Heating lamp
Temp. alarm lamp
Heater alarm lamp
Temp. control and blower operation lamp
Measured value indication lamp
Set value indication lamp
Temperature controller/blower RUN/STOP key
DOWN key
UP key
SET key
-21-
Over temperature protector
Actuation lamp
Overheating protector
actuation reset switch