4-4 Connection of utility
Check the voltage, phase and
capacity of power and correctly
connect the unit.
Wrong power connection may cause fire or
an incident of electrical shock.
Do not use a branch socket or
table tap.
Excess current may burn the cable or cause
fire.
Never connect a ground wire to the gas piping
or water piping to prevent an incident of
electrical shock.
Connect a ground wire correctly.
Warning
Warning
Warning
1)Check the voltage, phase and capacity of power to
which the product model is connected.
Power to which the product is connected is as follows.
2)No power plug is supplied.
Directly connect to the distribution board.
*Do not connect power yet here.
Connect the ground wire of the power cord to the
ground.
Product model
Power necessary for
connection
Power
Capacity
SD-1010
AC220Vsingle
phase
30A
-10-
If a ground wire is not connected, an incident of
electrical shock may occur.
Be sure to connect a ground wire.
Warning
Ground (green)
Power (white, black)
To AC220V single phase power