5-1 Preparation
Operation
5
-11-
Handle glass wares carefully.
Handle glass wares carefully not to break down.
Caution
5-1-1 Attaching evaporating tube
1) Place the rectification plate to the evaporating tube.
*There is no restraint etc. on direction of the current
plate.
2) Install the evaporating tube on the main unit.
Match the position of engraved mark
○
on the front
of the evaporating tube with the position of
engraved mark
○
on the evaporating tube
mounting portion of the main unit and insert the
evaporating tube into the main unit.
*During installation, you may put the evaporating
tube temporarily on the evaporating tube
receiving cradle of the main unit.
**When insertion becomes heavy (tight), replace O
ring inside the main unit.
3) After insertion into the main unit, rotate the
evaporating tube leftward at 45º.
A click on the evaporating tube mounting portion is
clutched and the evaporating tube is temporarily
fastened on the main unit.
4)Tighten the lock knob with the hand.
The evaporating tube is fastened on the main unit.
Rectification board
Evaporator
Evaporating tube
mounting portion
Engraved mark
○
Evaporator
Evaporating tube
receiving cradle
①
Rotate
Lock knob
Insert
Installation
②
Fasten
Code No.
Part name
Remarks
180210
O ring containing Si oil G175