68
WT.050.810.040.DE.IM.0817
Retrofit kits, spares and accessories
DEPOLOX
®
5 E flow cell
6.
13
Install the SiDiSens LF module on the left side in the flow cell
and secure with the enclosed plastic self-tapping screws
(W2T807965 - 10 mm) in position M - Figure 3 (mounting
recess).
Please note
Prior to the installation of the SiDiSens module, please ensure that
you use the correct plastic self-tapping screws. Use only the
plastic self-tapping screws A2 (W2T807965 - 10 mm) when
installing the SiDiSens module. Please distinguish between the
plastic self-tapping screws for installing the flow cell and those
used for the SiDiSens module. The screws should not be
confused.
14
Plug in the CAN sensor cable of the SiDiSens LF module at
position M - Figure 3. The M12 terminating resistor must be
screwed on CAN socket (N - Figure 4).
Figure 3 DEPOLOX
®
5 E flow cell, non-pressurized (cover
removed)
J
CAN sensor cable connection
K
SiDiSens LF connection
L
Membrane sensor connection
M Mounting recess
Plastic self-tapping screw (10 mm)
J
L
M
K
Summary of Contents for WALLACE & TIERNAN DEPOLOX 5 E
Page 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Page 8: ...8 WT 050 810 040 DE IM 0817 Introduction DEPOLOX 5 E flow cell 1...
Page 82: ...82 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 83: ...DEPOLOX 5 E flow cell Declarations and certificates 7 83 7 2 Certificate of CSA...
Page 84: ...84 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 85: ...DEPOLOX 5 E flow cell Declarations and certificates 7 85...
Page 86: ...86 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 90: ...90 WT 050 810 040 DE IM 0817 Index DEPOLOX 5 E flow cell 8...