DEPOLOX
®
5 E flow cell
Installation
4.
49
4.4.10
Removing the felt ring
To keep the diaphragm moist and prevent crystallization in the
electrolyte solution there is a moist felt washer in the gap between
electrolyte tank and electrode when the unit is in storage.
Attention!
Before initial startup, the felt washer for moistening the mem-
branes must be removed.
Proceed as follows:
1
Extract the knurled nut (A) from the cover of the electrolyte
tank.
2
Remove the entire electrode unit from the cell body from
below.
3
Remove the felt washer (B) between electrolyte tank and elec-
trode.
4
Refit the entire electrode unit in the cell body. With the locating
pin, observe the exact position in the cell body.
5
Push the entire electrode unit upwards.
6
Refit the knurled nut on the electrolyte tank.
Figure 7
Cross-section of cell body, non-pressurized version
(example knurled nut)
A
Knurled nut
B
Felt washer
A
B
Summary of Contents for WALLACE & TIERNAN DEPOLOX 5 E
Page 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Page 8: ...8 WT 050 810 040 DE IM 0817 Introduction DEPOLOX 5 E flow cell 1...
Page 82: ...82 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 83: ...DEPOLOX 5 E flow cell Declarations and certificates 7 83 7 2 Certificate of CSA...
Page 84: ...84 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 85: ...DEPOLOX 5 E flow cell Declarations and certificates 7 85...
Page 86: ...86 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 90: ...90 WT 050 810 040 DE IM 0817 Index DEPOLOX 5 E flow cell 8...