DEPOLOX
®
5 E flow cell
Installation
4.
47
4.4.8
Connecting the sample water outlet
Please note
Do not install water carrying lines made of copper. These would
falsify the measurements.
Sample water outlet on non-pressurized version
Proceed as follows:
1
On the non-pressurized version, no back-pressure is permitted
in the cell body.
2
The sample water outlet must be open. We recommend using
a funnel above the outlet.
3
The sample water outlet must be laid in such a way as to pre-
vent siphoning.
Figure 4
Cross-section of DEPOLOX
®
5 E flow cell, non-
pressurized
A
Sample water inlet with shut-off ball valve
B
Sample taking unit (drain on the drain screw)
C Sample water outlet
Sample water outlet on pressurized version
Proceed as follows:
1
On the pressurized version, a maximum back-pressure of
1.5 bar is permitted on the sample water outlet.
2
Ensure that the drain screw (sample taking unit) is always
closed.
A
B C
Summary of Contents for WALLACE & TIERNAN DEPOLOX 5 E
Page 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Page 8: ...8 WT 050 810 040 DE IM 0817 Introduction DEPOLOX 5 E flow cell 1...
Page 82: ...82 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 83: ...DEPOLOX 5 E flow cell Declarations and certificates 7 83 7 2 Certificate of CSA...
Page 84: ...84 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 85: ...DEPOLOX 5 E flow cell Declarations and certificates 7 85...
Page 86: ...86 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 90: ...90 WT 050 810 040 DE IM 0817 Index DEPOLOX 5 E flow cell 8...