62
WT.050.810.040.DE.IM.0817
Maintenance
DEPOLOX
®
5 E flow cell
5.
29
Fill in cleaning sand. See chapter 4.4.12 “Filling electrode
cleaning sand”.
30
Insert or screw on the sensor and connect with DFMe
electronics module.
31
Refit and engage the housing cover of DEPOLOX
®
5 E flow
cell.
32
Open the shut-off ball valve (D) at the sample water inlet.
33
On the pressurized version, open the shut-off ball valve on the
sample water outlet.
34
The cell body (A) refills with sample water.
35
Re-establish the power supply.
Please note
After topping up or replacing cleaning sand, the electrode current
may increase slightly for about three hours. Do not calibrate during
this time. Recalibrate after each cell sand replacement. The cali-
bration must be checked after one day.
Please note
After 2 to 3 hours running-in time, perform a chlorine calibration. If
necessary, repeat the chlorine calibration after 24 hours. See
section Calibration in instruction manual “700 P electronics
module” of “700 M electronics module”. An initially rotating air
bubble at the bottom of the cell body does not affect the
measurement.
Summary of Contents for WALLACE & TIERNAN DEPOLOX 5 E
Page 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Page 8: ...8 WT 050 810 040 DE IM 0817 Introduction DEPOLOX 5 E flow cell 1...
Page 82: ...82 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 83: ...DEPOLOX 5 E flow cell Declarations and certificates 7 83 7 2 Certificate of CSA...
Page 84: ...84 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 85: ...DEPOLOX 5 E flow cell Declarations and certificates 7 85...
Page 86: ...86 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 90: ...90 WT 050 810 040 DE IM 0817 Index DEPOLOX 5 E flow cell 8...