Instruction Manual
pH 1100/ 2100
54
10.5.8
Memory clear setup mode P4.6
Clears all pH, mV and Ion (pH2100 only) memory points stored
in memory.
1. Enter meter setup mode as described in steps 1-4 in
Section
10.5.1 Entering meter setup mode
(page 50).
2. Use
∆
/MI
key to scroll to program
Mem Clear Setup: P4.6
.
See
Figure 181
.
3. Press
ENTER
to enter program
Mem Clear Setup: P4.6
.
4. Press
>/CAL
key to toggle between
Memory Clear Yes
and
No
.
The selection will flash. See
Figure 182
.
5. Press
ENTER
to confirm selection.
•
If you select
Yes
, the screen flashes
Memory being
Cleared, Please wait...
See
Figure 183
. The meter then
returns to the setup mode.
•
If you select
No
, the meter returns to the setup mode.
Press
∆
/MI
to continue to the next setup program, or
press
∇
/MR
to go to the previous program. Press
EXIT
to
return to measurement mode.
10.5.9
Meter reset setup mode P4.7
Resets the meter to factory default settings:
• pH/mV/Ion* calibration data is erased
• Manual temperature compensation is set to 25.0°C
• pH/mV/Ion* measurement alarm is off
• pH/mV/Ion* calibration due alarm is off
• Stability indicator is turned on with low stability setting
• Data logging and memory recall has manual setting
• Backlight is on with 2.0 minute duration
• Memory is cleared
*Ion features on the CyberScan pH2100 meter only
1. Enter meter setup mode as described in steps 1-4 in
Section
10.5.1 Entering meter setup mode
(page 50).
2. Use
∆
/MI
key to scroll to program
Meter Reset: P4.7
.
See
Figure 184
.
3. Press
ENTER
to enter program
Meter Reset: P4.7
.
4. Press
>/CAL
key to toggle between
Meter Reset Yes
and
No
.
The selection will flash. See
Figure 185
.
5. Press
ENTER
to confirm selection.
•
If you selected
Yes
, the screen flashes
Meter being
Reset, Please wait...
See
Figure 186
. The meter then
returns to measurement mode.
•
If you select
No
, the screen flashes
Reset Aborted.
See
Figure 187
. The meter then returns to the measurement
mode.
Memory being Cleared
Please wait. . .
Memory Clear:Yes
No
Select & Press ENTER
Mem Clear Setup :P4.6
Press ENT. to Select
Figure 181
Figure 182
Figure 183
Meter being Reset
Please wait. . .
Meter Reset: Yes
No
Select & Press ENTER
Meter Reset :P4.7
Press ENT. to Select
Figure 184
Figure 185
Figure 186
Figure 187
Reset Aborted