Instruction Manual
pH 1100/ 2100
45
10.4.4
Ion calibration due alarm setup program P3.2
Use this program to set a reminder when the next calibration is due.
Choose from 1 to 365 days from last calibration.
1. Enter ion setup mode as described in steps 1-4 in
Section 10.4.1
Entering ion setup mode
(page 43).
2. Use
∆
/MI
key to scroll to program
IonCalDue Alarm: P3.2
.
See
Figure 138
.
3. Press
ENTER
to enter program
IonCalDue Alarm: P3.2
.
4. Press
>/CAL
key to toggle between
Cal Alarm Due: On
and
Off
.
The selection will flash. See
Figure 139
.
5. Press
ENTER
to confirm selection.
•
If you selected
On
, continue to step 6.
•
If you select
Off
, pressing
ENTER
returns to the setup mode.
Press
∆
/MI
key to continue to the next setup program, or press
∇
/MR
to go to the previous program. Press
EXIT
to return to mV
measurement mode.
6. If you selected On in step 5, the display reads
Key in Cal Due
Period (1-365 days) = 1
. Key number of days using the numeric
keypad. If you make a mistake, use
∆
/MI
or
∇
/MR
keys to
highlight mistake, then re-enter. See
Figure 140
.
7. Press
ENTER
to confirm number of days to next calibration;
screen flashes
Sys Updating, Please wait…
and returns to the
setup mode. Press
∆
/MI
key to continue to the next setup
program, or press
∇
/MR
to go to the previous program. Press
EXIT
to return to mV measurement mode.
Keyin Cal Due Period
(1-365 days) = 1
Cal Alarm Due: On Off
Select & Press ENTER
IonCalDue Alarm :P3.2
Press ENT. to Select
Figure 138
Figure 139
Figure 140