Instruction Manual
pH 1100/ 2100
46
10.4.5
Stability setup program P3.3
Use this program to turn on or off the ion stability indicator and
to set level of stability (low, medium or high).
Stability definitions
High stability:
Reading is stabilized quickly, but reading is less accurate.
Medium stability:
Reading stability is averaged between high and low stability.
Low stability:
Reading is stabilized slowly, but guarantees high accuracy.
1. Enter ion setup mode as described in steps 1-4 in
Section
10.4.1 Entering ion setup mode
(page 43).
2. Use
∆
/MI
key to scroll to program
Stability Setup: P3.3
See
Figure 141
.
3. Press
ENTER
to enter program
Stability Setup: P3.3
4. Press
>/CAL
key to toggle between
Ion Stable Criteria: On
and
Off
. The selection will flash. See
Figure 142
.
5. Press
ENTER
to confirm selection.
•
If you selected
On
, continue to step 6.
•
If you select
Off
, pressing
ENTER
returns to the setup
mode.
Press
∆
/MI
key to continue to the next setup program, or
Press
∇
/MR
to go to the previous program. Press
EXIT
to
return to the ion measurement mode.
6. If you selected
On
in step 5, the display now reads Degree of
Stability: Low Medium High
. . Press
>/CAL
to toggle between
Low, Medium
and
High
. The selection will flash. See
Figure
143
.
7. Press
ENTER
to confirm selection and return to the setup
mode. Press
∆
/MI
key to continue to the next setup program,
or press
∇
/MR
to go to the previous program. Press
EXIT
to
return to the ion measurement mode.
Ion Stable Criteria
On Off
Degree of Stability
Low Medium High
Stability Setup : P3.3
Press ENT. to Select
Figure 141
Figure 142
Figure 143