Instruction Manual
pH 1100/ 2100
32
10.2.5
pH Measurement Alarm Setup Program P1.3
Use this program to select alarm for low and high pH
measurements.
1. Enter
pH/Temperature
setup
mode as described in steps 1-4
in
10.2.1 Entering pH/temperature setup mode
(page 30).
2. Use
∆
/MI
to scroll to program
pH Meas. Alarm: P1.3
. See
Figure 87
.
3. Press
ENTER
to enter program
pH Meas. Alarm: P1.3
.
4. Press
>/CAL
to toggle between pH Alarm
On
and
Off
. The
selection will flash. See
Figure 88
.
5. Press
ENTER
to confirm selection.
•
If you selected
On
continue to step 6
•
If you select
Off,
pressing
ENTER
returns to the setup
mode.
Press
∆
/MI
key to continue to the next setup program, or press
∇
/MR
to go to the previous program
.
Press EXIT to return to
the pH measurement mode.
6. If
you
selected
On
, the display now reads
Key in Low Alarm
pH Set Point
. Key in the low pH alarm set point using the
numeric keypad. If you make a mistake, use the
∆
/MI
or
∇
/MR
keys to highlight mistake, then re-enter. See
Figure 89
. Press
ENTER
to confirm keyed in value.
7. The display now reads
Key in High Alarm pH Set Point
. Key in
the high pH alarm set point using the numeric keypad. The
high alarm must be set at least 1.0 pH greater than the low
alarm. If you make a mistake, use the
∆
/MI
or
∇
/MR
keys to
highlight mistake, then re-enter. See
Figure 90
.
8. Press
ENTER
to confirm high pH set point value and return to
the setup mode.
Press
∆
/MI
key to continue to the next setup program, or
press
∇
/MR
to go to the previous program
.
Press
EXIT
to
return to the pH measurement mode.
Figure 89
Figure 90
Key In Low Alarm
pH Set Point:2 . 000
Key In High Alarm
pH Set Point:14 . 000
pH Meas. Alarm :P1.3
Press ENT. to Select
Figure 87
Figure 88
pH Alarm On
Select & Press ENTER
Off