10.3.3
mV calibration due alarm setup program P2.1
Use this program to set a reminder when the next calibration is
due. Choose from 1 to 365 days from last calibration.
1. Enter mV setup mode as described in steps 1-4 in
Section
10.3.1 Entering mV setup mode
(page 37).
2. Use
∆
/MI
key to scroll to program
mVCalDue Alarm: P2.1
.
See
Figure 116
.
3. Press
ENTER
to enter program
mVCalDue Alarm: P2.1
.
4. Press
>/CAL
to toggle between Cal Alarm Due:
On
and
Off
.
The selection will flash. See
Figure 117
.
5. Press
ENTER
to confirm selection.
•
If you selected
On,
continue to step 6.
•
If you select
Off,
pressing
ENTER
returns to the setup
mode.
Press
∆
/MI
to continue to the next setup program, or press
∇
/MR
to go to the previous program
.
Press
EXIT
to return to
mV measurement mode.
6. If you selected
On
in step 5, the display reads
Key in Cal
Due Period (1-365 days) =
. Key number of days using the
numeric keypad. If you make a mistake, use the
∆
/MI
or
∇
/MR
keys to highlight mistake, then re-enter.
See
Figure 118
.
7. Press
ENTER
to confirm number of days to next calibration,
screen flashes
Sys Updating, Please wait...
and returns to
the setup mode.
Press
∆
/MI
to continue to the next setup program, or press
∇
/MR
to go to the previous program
.
Press
EXIT
to return to
mV measurement mode.
Keyin Cal Due Period
(1-365 days) = 1
Cal Alarm Due On Off
Select & Press ENTER
mVCalDue Alarm :P2.1
Press ENT. to Select
Figure 116
Figure 117
Figure 118