Instruction Manual
pH 1100/ 2100
12
3.1.2
Starting pH calibration
1. Turn the meter
ON
. If the meter is not in the pH measurement
mode, press
EXIT
and press
>/CAL
to select pH/Temp. The
mode selected will flash. See
Figure 13
. Press
ENTER
to
confirm selection. The meter is now in pH measurement mode.
See
Figure 14
.
2. Rinse the electrode in deionized water or rinse solution. If using
the ATC function with a separate temperature probe, rinse the
temperature probe as well.
Do not wipe the pH electrode or
temperature probe dry.
Wiping the probes causes static, and
will create calibration and measurement instability.
3. Select pH buffer and pour into a clean container.
4.
Dip the electrode and ATC probe into the calibration buffer. The
end of the probes must be completely immersed into the
sample. Stir probes gently to create a homogeneous sample.
5. Press
>/CAL
to enter the pH calibration mode. The display
flashes
Cal Mode Activated, Please wait..
..
See
Figure 15
. The meter is now in pH calibration mode.
The meter automatically recognizes the buffers in the set you have selected in the
SETUP
mode. If you wish to change
the buffer set selected, see
Section 10.2.2, Program pH Buffer Setup: P1.0
in the
pH/Temp
setup mode (page 30).
If your meter is set up for standard buffers (USA, NIST, Bf1, Bf2), proceed to
Section 3.1.3
:
Standard pH buffer
calibration
(page 13). If your meter is set up for custom buffers, proceed to
Section 3.1.4
:
Custom pH buffer calibration
(page 14).
Figure 13
Figure 14
Figure 15
pH/Temp mV Ion
Select & Press ENTER
7.179pH 25.3 C ATC
Jan 18 2002 08:46:25
°
Cal Mode Activated
Please wait. . .