Ion mode setup program P3.4
Use this program to select high or low ion resolution mode. High
resolution has ±500 mV value, low has ±1850 mV value.
1. Enter ion setup mode as described in steps 1-4 in
Section
10.4.1 Entering ion setup mode
(page 43).
2. Use
∆
/MI
key to scroll to program
Ion Mode Setup: P3.4
. See
Figure 144
.
3. Press
ENTER
to enter program
Ion Mode Setup: P3.4
.
4. Press
>/CAL
to toggle between
High
or
Low
. The unit
selected will flash. See
Figure 145
.
5. Press
ENTER
to confirm selection and return to the setup
mode.
6. Press
∆
/MI
key to continue to the next setup program, or
press
∇
/MR
to go to the previous program. Press
EXIT
to
return to the ion measurement mode.
Resolution: High Low
Select & Press ENTER
Ion Mode Setup : P3.4
Press ENT. to Select