Instruction Manual
pH 1100/ 2100
36
10.2.9
pH calibration data reset setup program P1.7
Use this program to erase all pH calibration data.
1. Enter pH/Temperature setup mode as described in steps 1-4 in
Section 10.2.1 Entering pH/temperature setup mode
(page 30).
2. Press
∆
/MI
to scroll to program
pH Cal Reset: P1.7
. See
Figure 105
.
3. Press
ENTER
to enter program
pH Cal Reset: P1.7
.
4. Press
>/CAL
to toggle between pH Reset:
Yes
and
No
. The
selection will flash. See
Figure 106
.
5. Press
ENTER
to confirm selection.
•
If you selected
Yes,
the screen flashes
pH being Reset,
Please wait...
See
Figure 107
. The meter then returns to
setup mode.
•
If you select
No,
the screen flashes
Reset Aborted
and
then returns to the setup mode. See
Figure 108
.
This is the last pH setup program. Press
∇
/MR
to go to the previous
program. Press
∆
/MI
to go back to the first pH setup program. Press
EXIT
to return to the pH measurement mode.
Reset Aborted
pH being Reset
Please wait. . .
pH Reset: Yes No
Select & Press ENTER
pH Cal Reset :P1.7
Press ENT. to Select
Figure 105
Figure 106
Figure 107
Figure 108