System setup sub-group allows you to configure general settings of the meter.
The settings are displayed in 6 pages. Press
NEXT-P (F2)
and
PREV-P (F1)
to
navigate through these pages.
Parameter
Description
Factory Default
STABLE
Indicator
ENABLE - The meter displays ‘Stable’ indicator in the
measurement screen as per the ‘STABLE CRITERIA’
defined below.
DISABLE – ‘Stable’ indicator does not appear.
ENABLE
Stability
Criteria
SLOW – The reading is stabilized slowly and exhibits
good repeatability
MEDIUM – Reading stability is averaged between
slow & fast stability
FAST – Reading is stabilized quickly at the cost of
repeatability.
(This parameter has no effect if ‘STABLE’ parameter
is disabled)
FAST
Auto Hold
ENABLE - The meter holds the reading in the
measurement screen, if the reading is ‘Stable’ for 5
seconds.
If this is enabled, ‘Response time’ appears in the
measurement screen, indicating the average response
time of the probe.
DISABLE – The reading is not held
(This parameter has no effect if ‘STABLE’ parameter
is disabled. The response time may not work if the
system time has not been set as described previously.
DISABLE
Tem. Display
from
Allows you to select temperature from pH/COND/DO
probes to display in multi measurement screen.
For PCD 650-pH/COND->DO
-
Display
setting:-
Allows you to select multi modes that you would like to
be displayed on the 1st and 2
nd
row of the
measurement screen after calibration.
1
st
Row : pH / Ion / mV / Conductivity / Salinity /
Resitivity / TDS / O
2
mg/L (ppm) / O
2
(%)
2
nd
Row : pH / Ion / mV / Conductivity / Salinity /
Resitivity / TDS / O
2
mg/L (ppm) / O
2
(%)
3
rd
Row : pH / Ion / mV / Conductivity / Salinity /
Resitivity / TDS / O
2
mg/L (ppm) / O
2
(%)
-
Note:
In order to activate the RESPONSE TIME function, you have to first
activate the STABLE and AUTO HOLD functions.
Summary of Contents for CD 650
Page 2: ......
Page 4: ......
Page 10: ...2 ...
Page 28: ...20 ...
Page 45: ...37 ...
Page 49: ...41 ...
Page 51: ...43 ...
Page 61: ...53 ...
Page 65: ...57 ...
Page 67: ...59 ...
Page 73: ...65 ...
Page 82: ...74 END OF CONDUCTIVITY SECTION ...
Page 84: ...76 ...
Page 92: ...84 ...
Page 98: ...90 ...
Page 100: ...92 ...
Page 106: ...98 ...
Page 108: ...100 ...
Page 126: ...118 ...